Development of up-converting phosphor technology-based lateral-flow assay for rapidly quantitative detection of hepatitis B surface antibody
- PMID: 19150709
- DOI: 10.1016/j.diagmicrobio.2008.10.020
Development of up-converting phosphor technology-based lateral-flow assay for rapidly quantitative detection of hepatitis B surface antibody
Abstract
An up-converting phosphor technology-based lateral-flow (UPT-LF) assay system was developed for rapid and quantitative detection of hepatitis B surface antibody (HBsAb). To evaluate its performance, we compared it with the Abbott Axsym AUSAB (ABBOTT Diagnostics Division, Wiesbaden, Germany) assay and conventional ELISA (Wantai Biological Pharmacy Enterprise, Beijing, China) using 13 standard positive sera and 306 clinical sera. In both laboratory evaluation and clinical application, UPT-LF assay showed the best sensitivity (99.19%) and detection agreement (97.43% for the adjusted agreement) with true results. The concordance rate between UPT-LF and ELISA, as shown by correlative regression analysis, was the highest (R(2)=0.6389), whereas that between UPT-LF and AUSAB was the lowest (R(2)=0.5702). In conclusion, UPT-LF assay for quantitative detection of HBsAb is sensitive and rapid, promising this new assay a bright future.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical