Intra-patient reproducibility of myocardial SPECT imaging with 201Tl

Affiliations
Comparative Study

Intra-patient reproducibility of myocardial SPECT imaging with 201Tl

Barbra E Backus et al. J Nucl Cardiol. 2009 Jan-Feb.

Abstract

Background: To define the physical and clinical reproducibility of (201)Tl myocardial perfusion SPECT (MPS), this study assesses the variation between two repeated rest (201)Tl MPS with repositioning only, with a two-hour time interval and with phantom measurements as a reference.

Methods: Three anthropomorphic thorax phantoms were filled with (201)Tl. For each phantom five repeated (201)Tl MPS were obtained. In addition, in 20 patients repeated (201)Tl rest-MPS and in 26 patients early and delayed (201)Tl rest-MPS were performed. Quantitative analysis was done using MunichHeart. Statistical methods were used to calculate variability. Visual analysis was performed by 2 independent observers.

Results: The average variation between repeated phantom MPS was 0.5% (95% confidence interval (CI): -0.4% to 1.4%). For patient scans this was -5.0% (95% CI: -2.5% to -7.5%) and between early and delayed (201)Tl MPS -15.5% (95% CI: -11.7% to -19.3%). Visual assessment revealed no clinical significant differences between rest (201)Tl and repeated or delayed (201)Tl MPS.

Conclusions: Repositioning in phantom (201)Tl MPS does not cause significant variation. Repeated (201)Tl MPS in patients shows 5.0% decrease of (201)Tl in 30 minutes, which increases to 15% during a two-hour time interval without quantitative or visual regional differences. This decrease indicates a time-related washout of (201)Tl, but does not change clinical diagnosis.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Eur J Nucl Med Mol Imaging. 2002 Aug;29(8):1063-71 - PubMed
    1. Eur J Nucl Med. 1998 Sep;25(9):1313-21 - PubMed
    1. Circulation. 1981 Sep;64(3):610-8 - PubMed
    1. Eur J Nucl Med Mol Imaging. 2003 Jan;30(1):25-31 - PubMed
    1. J Nucl Cardiol. 2005 Jul-Aug;12(4):401-9 - PubMed

Publication types

LinkOut - more resources