Intra-patient reproducibility of myocardial SPECT imaging with 201Tl
- PMID: 19152134
- DOI: 10.1007/s12350-008-9011-7
Intra-patient reproducibility of myocardial SPECT imaging with 201Tl
Abstract
Background: To define the physical and clinical reproducibility of (201)Tl myocardial perfusion SPECT (MPS), this study assesses the variation between two repeated rest (201)Tl MPS with repositioning only, with a two-hour time interval and with phantom measurements as a reference.
Methods: Three anthropomorphic thorax phantoms were filled with (201)Tl. For each phantom five repeated (201)Tl MPS were obtained. In addition, in 20 patients repeated (201)Tl rest-MPS and in 26 patients early and delayed (201)Tl rest-MPS were performed. Quantitative analysis was done using MunichHeart. Statistical methods were used to calculate variability. Visual analysis was performed by 2 independent observers.
Results: The average variation between repeated phantom MPS was 0.5% (95% confidence interval (CI): -0.4% to 1.4%). For patient scans this was -5.0% (95% CI: -2.5% to -7.5%) and between early and delayed (201)Tl MPS -15.5% (95% CI: -11.7% to -19.3%). Visual assessment revealed no clinical significant differences between rest (201)Tl and repeated or delayed (201)Tl MPS.
Conclusions: Repositioning in phantom (201)Tl MPS does not cause significant variation. Repeated (201)Tl MPS in patients shows 5.0% decrease of (201)Tl in 30 minutes, which increases to 15% during a two-hour time interval without quantitative or visual regional differences. This decrease indicates a time-related washout of (201)Tl, but does not change clinical diagnosis.
Similar articles
-
Imaging technology for myocardial perfusion single-photon emission computed tomography 2018 in Japan.Jpn J Radiol. 2020 Mar;38(3):274-282. doi: 10.1007/s11604-019-00915-2. Epub 2020 Jan 9. Jpn J Radiol. 2020. PMID: 31919636
-
Degradation of myocardial perfusion SPECT images caused by contaminants in thallous (201Tl) chloride.Eur J Nucl Med Mol Imaging. 2008 May;35(5):922-32. doi: 10.1007/s00259-008-0719-z. Epub 2008 Jan 25. Eur J Nucl Med Mol Imaging. 2008. PMID: 18219482
-
Scatter correction on its own increases image contrast in Tl-201 myocardium perfusion scintigraphy, but does it also improve diagnostic accuracy?Ann Nucl Med. 2003 Dec;17(8):725-31. doi: 10.1007/BF02984983. Ann Nucl Med. 2003. PMID: 14971620 Clinical Trial.
-
Automatic detection and size quantification of infarcts by myocardial perfusion SPECT: clinical validation by delayed-enhancement MRI.J Nucl Med. 2005 May;46(5):728-35. J Nucl Med. 2005. PMID: 15872343 Clinical Trial.
-
Effect of time and exercise on the clearance rate of (201)Tl in normal and ischemic myocardium.Nucl Med Commun. 2010 Jun;31(6):539-46. Nucl Med Commun. 2010. PMID: 20429095 Clinical Trial.
Cited by
-
A study of SPECT/CT camera stability for quantitative imaging.EJNMMI Phys. 2016 Dec;3(1):14. doi: 10.1186/s40658-016-0150-7. Epub 2016 Jul 29. EJNMMI Phys. 2016. PMID: 27473290 Free PMC article.
-
Accuracy of washout rate analysis for thallium-201 single-photon emission computed tomography myocardial perfusion imaging using cadmium zinc telluride detectors: A phantom study.JRSM Cardiovasc Dis. 2020 Jan 13;9:2048004019900600. doi: 10.1177/2048004019900600. eCollection 2020 Jan-Dec. JRSM Cardiovasc Dis. 2020. PMID: 31976070 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources