Detection of breast cancer with full-field digital mammography and computer-aided detection
- PMID: 19155392
- DOI: 10.2214/AJR.07.3884
Detection of breast cancer with full-field digital mammography and computer-aided detection
Abstract
Objective: The purpose of this study was to evaluate computer-aided detection (CAD) performance with full-field digital mammography (FFDM).
Materials and methods: CAD (Second Look, version 7.2) was used to evaluate 123 cases of breast cancer detected with FFDM (Senographe DS). Retrospectively, CAD sensitivity was assessed using breast density, mammographic presentation, histopathology results, and lesion size. To determine the case-based false-positive rate, patients with four standard views per case were included in the study group. Eighteen unilateral mammography examinations with nonstandard views were excluded, resulting in a sample of 105 bilateral cases.
Results: CAD detected 115 (94%) of 123 cancer cases: six of six (100%) in fatty breasts, 63 of 66 (95%) in breasts containing scattered fibroglandular densities, 43 of 46 (93%) in heterogeneously dense breasts, and three of five (60%) in extremely dense breasts. CAD detected 93% (41/44) of cancers manifesting as calcifications, 92% (57/62) as masses, and 100% (17/17) as mixed masses and calcifications. CAD detected 94% of the invasive ductal carcinomas (n = 63), 100% of the invasive lobular carcinomas (n = 7), 91% of the other invasive carcinomas (n = 11), and 93% of the ductal carcinomas in situ (n = 42). CAD sensitivity for cancers 1-10 mm (n = 55) was 89%; 11-20 mm (n = 37), 97%; 21-30 mm (n = 16), 100%; and larger than 30 mm (n = 15), 93%. The CAD false-positive rate was 2.3 marks per four-image case.
Conclusion: CAD with FFDM showed a high sensitivity in identifying cancers manifesting as calcifications and masses. Sensitivity was maintained in cancers with lower mammographic sensitivity, including invasive lobular carcinomas and small neoplasms (1-20 mm). CAD with FFDM should be effective in assisting radiologists with earlier detection of breast cancer. Future studies are needed to assess CAD accuracy in larger populations.
Similar articles
-
Performance of computer-aided detection applied to full-field digital mammography in detection of breast cancers.Eur J Radiol. 2011 Mar;77(3):457-61. doi: 10.1016/j.ejrad.2009.08.024. Epub 2009 Oct 28. Eur J Radiol. 2011. PMID: 19875260
-
[Impact of breast density on computer-aided detection (CAD) of breast cancer].Zhonghua Zhong Liu Za Zhi. 2012 May;34(5):360-3. doi: 10. 3760/cma.j.issn.0253-3766.2012.05.009. Zhonghua Zhong Liu Za Zhi. 2012. PMID: 22883457 Chinese.
-
Screening mammography-detected cancers: sensitivity of a computer-aided detection system applied to full-field digital mammograms.Radiology. 2007 Jul;244(1):104-11. doi: 10.1148/radiol.2441060756. Epub 2007 May 16. Radiology. 2007. PMID: 17507722
-
Studies comparing screen-film mammography and full-field digital mammography in breast cancer screening: updated review.Acta Radiol. 2009 Jan;50(1):3-14. doi: 10.1080/02841850802563269. Acta Radiol. 2009. PMID: 19037825 Review.
-
The efficacy of using computer-aided detection (CAD) for detection of breast cancer in mammography screening: a systematic review.Acta Radiol. 2019 Jan;60(1):13-18. doi: 10.1177/0284185118770917. Epub 2018 Apr 17. Acta Radiol. 2019. PMID: 29665706
Cited by
-
A new approach to develop computer-aided detection schemes of digital mammograms.Phys Med Biol. 2015 Jun 7;60(11):4413-27. doi: 10.1088/0031-9155/60/11/4413. Epub 2015 May 18. Phys Med Biol. 2015. PMID: 25984710 Free PMC article.
-
Breast cancer screening: an evidence-based update.Med Clin North Am. 2015 May;99(3):451-68. doi: 10.1016/j.mcna.2015.01.002. Epub 2015 Mar 5. Med Clin North Am. 2015. PMID: 25841594 Free PMC article. Review.
-
Computer-aided detection; the effect of training databases on detection of subtle breast masses.Acad Radiol. 2010 Nov;17(11):1401-8. doi: 10.1016/j.acra.2010.06.009. Epub 2010 Jul 22. Acad Radiol. 2010. PMID: 20650667 Free PMC article.
-
Computer-aided detection of breast masses depicted on full-field digital mammograms: a performance assessment.Br J Radiol. 2012 Jun;85(1014):e153-61. doi: 10.1259/bjr/51461617. Epub 2011 Feb 22. Br J Radiol. 2012. PMID: 21343322 Free PMC article.
-
CAD May Not be Necessary for Microcalcifications in the Digital era, CAD May Benefit Radiologists for Masses.J Clin Imaging Sci. 2012;2:45. doi: 10.4103/2156-7514.99179. Epub 2012 Jul 28. J Clin Imaging Sci. 2012. PMID: 22919559 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous