False-negative diagnoses at stereotactic vacuum-assisted needle breast biopsy: long-term follow-up of 1,280 lesions and review of the literature
- PMID: 19155393
- DOI: 10.2214/AJR.08.1127
False-negative diagnoses at stereotactic vacuum-assisted needle breast biopsy: long-term follow-up of 1,280 lesions and review of the literature
Abstract
Objective: The purpose of this study was to determine retrospectively the frequency and causes of false-negative diagnoses of lesions evaluated with percutaneous stereotactic biopsy performed with vacuum-assisted probes with the patient prone on a biopsy table.
Materials and methods: The records of 1,152 women (median age, 55 years; range, 31-94 years) with 1,280 lesions consecutively biopsied with 14-gauge (n = 159) or 11-gauge (n = 1,121) vacuum probes were reviewed retrospectively. The histologic diagnoses were malignant (n = 489), high-risk (n = 117), and benign (n = 674) lesions. Benign lesions were evaluated with repeated biopsy (n = 46) or mammographic follow-up for 24 months or longer (n = 506) or less than 24 months (n = 57). In 65 cases (10% of all benign lesions), follow-up was not performed. Fisher's exact tests were used to correlate patient, mammographic, and biopsy variables with false-negative diagnoses.
Results: Of 508 lesions with a final diagnosis of malignancy, the initial needle biopsy diagnosis was malignant (n = 489), high-risk (n = 14), or benign (n = 5) lesion. The five false-negative diagnoses were related to gauge of the biopsy probe and specimen radiographic findings regarding calcifications but were not related to lesion type. False-negative diagnoses were made in 4.4% (three of 68) of cases in which 14-gauge vacuum biopsy was performed and in 0.45% (two of 440) of cases in which 11-gauge vacuum biopsy was performed (p = 0.019). False-negative diagnoses were made in 25% (one of four) of cases in which specimen radiographs showed no calcifications and 0.67% (two of 300) of cases in which they did show calcifications (p = 0.0390). False-negative diagnoses were made in 1.2% (three of 248) of cases of calcification lesions and 0.8% (two of 260) of cases of mass lesions (p = 0.679).
Conclusion: False-negative findings at stereotactic biopsy were least common with 11-gauge probes and were similar in frequency for calcifications and masses.
Similar articles
-
Stereotactic 11-gauge vacuum-assisted breast biopsy: a validation study.AJR Am J Roentgenol. 2002 Dec;179(6):1503-7. doi: 10.2214/ajr.179.6.1791503. AJR Am J Roentgenol. 2002. PMID: 12438044
-
MR imaging-guided 9-gauge vacuum-assisted core-needle breast biopsy: initial experience.Radiology. 2006 Jan;238(1):54-61. doi: 10.1148/radiol.2381050050. Epub 2005 Nov 22. Radiology. 2006. PMID: 16304093
-
Lobular neoplasia at 11-gauge vacuum-assisted stereotactic biopsy: correlation with surgical excisional biopsy and mammographic follow-up.AJR Am J Roentgenol. 2006 Oct;187(4):949-54. doi: 10.2214/AJR.05.0710. AJR Am J Roentgenol. 2006. PMID: 16985141
-
Present state of and problems with core needle biopsy for non-palpable breast lesions.Breast Cancer. 2006;13(1):32-7. doi: 10.2325/jbcs.13.32. Breast Cancer. 2006. PMID: 16518060 Review.
-
False-positive reports in fine needle biopsy of breast lesions.Pathology. 2001 Nov;33(4):428-36. doi: 10.1080/00313020120083142. Pathology. 2001. PMID: 11827408 Review.
Cited by
-
Stereotactic breast biopsies: Radiological-pathological concordance in a South African referral unit.SA J Radiol. 2022 Aug 26;26(1):2463. doi: 10.4102/sajr.v26i1.2463. eCollection 2022. SA J Radiol. 2022. PMID: 36093214 Free PMC article.
-
The current role of vacuum assisted breast biopsy system in breast disease.J Breast Cancer. 2011 Mar;14(1):1-7. doi: 10.4048/jbc.2011.14.1.1. Epub 2011 Mar 31. J Breast Cancer. 2011. PMID: 21847387 Free PMC article.
-
Results of Short-Term Follow-Up in BI-RADS 3 and 4a Breast Lesions with a Histological Diagnosis of Fibroadenoma at Percutaneous Needle Biopsy.Breast Care (Basel). 2017 Sep;12(4):238-242. doi: 10.1159/000477536. Epub 2017 Aug 29. Breast Care (Basel). 2017. PMID: 29070987 Free PMC article.
-
Values of pathological analysis of lost tissue fragments in the vacuum canister during a vacuum-assisted stereotactic biopsy of the breast.Br J Radiol. 2013 May;86(1025):20120270. doi: 10.1259/bjr.20120270. Epub 2013 Mar 21. Br J Radiol. 2013. PMID: 23520227 Free PMC article.
-
Short interval follow-up after a benign concordant MR-guided vacuum assisted breast biopsy--is it worthwhile?Eur Radiol. 2014 Jun;24(6):1176-85. doi: 10.1007/s00330-014-3125-x. Epub 2014 Mar 14. Eur Radiol. 2014. PMID: 24623366
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical