Consistency of apex locator function: a clinical study
- PMID: 19166768
- DOI: 10.1016/j.joen.2008.10.017
Consistency of apex locator function: a clinical study
Abstract
The consistency of apex locators was determined by calculating the dysfunction frequency. Electronic working length (EWL) was determined in 507 patients requiring endodontic treatment. Different clinical parameters were recorded including tooth vitality, presence of obliteration, and metallic restoration. Two apex locators were used (Root ZX [Morita, Tokyo, Japan] and Raypex5 [VDW, Munich, Germany]). Apex locator performance was considered "consistent" when the scale bars were stable and moved only in correspondence to the movement of file in the root canal. A working length radiograph with files set to the EWL was performed. EWL were considered "acceptable" when the file tip was located 0 to 2 mm short of the radiographic apex. The function of apex locators was consistent in 85% of the patients (429/507 [99% confidence interval, 80-88]). The inconsistent measurements were strongly associated with partially or totally obliterated root canals (p < 0.0001). Radiographically, 97% of consistent measurements were "acceptable."
Similar articles
-
Influence of preflaring on the accuracy of length determination with four electronic apex locators.J Endod. 2009 Sep;35(9):1300-2. doi: 10.1016/j.joen.2009.05.030. J Endod. 2009. PMID: 19720236
-
A comparison between two electronic apex locators: an ex vivo investigation.Int Endod J. 2007 May;40(5):362-73. doi: 10.1111/j.1365-2591.2006.01229.x. Epub 2007 Feb 19. Int Endod J. 2007. PMID: 17309741
-
Combination of apex locator and endodontic motor for continuous length control during root canal treatment.Int Endod J. 2009 Apr;42(4):368-74. doi: 10.1111/j.1365-2591.2008.01535.x. Epub 2009 Feb 7. Int Endod J. 2009. PMID: 19220512
-
Electronic apex locators.Int Endod J. 2004 Jul;37(7):425-37. doi: 10.1111/j.1365-2591.2004.00835.x. Int Endod J. 2004. PMID: 15189431 Review.
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
Cited by
-
Reliability and accuracy of dental MRI for measuring root canal length of incisors and canines: a clinical pilot study.Sci Rep. 2022 Aug 18;12(1):14068. doi: 10.1038/s41598-022-17889-3. Sci Rep. 2022. PMID: 35982139 Free PMC article.
-
Comparison of Tentative Radiographic Working Length with and without grid Versus Electronic Apex Locator.Contemp Clin Dent. 2018 Jan-Mar;9(1):88-91. doi: 10.4103/ccd.ccd_790_17. Contemp Clin Dent. 2018. PMID: 29599591 Free PMC article.
-
A comparative clinical evaluation of accuracy of six apex locators with intraoral periapical radiograph in multirooted teeth: An in vivo study.J Conserv Dent. 2017 Jul-Aug;20(4):264-268. doi: 10.4103/JCD.JCD_131_17. J Conserv Dent. 2017. PMID: 29259365 Free PMC article.
-
Evaluation of the effects of different file systems and apical functions of integrated endodontic motors on debris extrusion: an ex vivo experimental study.Restor Dent Endod. 2025 May;50(2):e14. doi: 10.5395/rde.2025.50.e14. Epub 2025 Apr 14. Restor Dent Endod. 2025. PMID: 40350563 Free PMC article.
-
Comparative evaluation of the accuracy of six different apex locators in working length determination of molars using intraoral periapical radiographs: An in vivo study.J Conserv Dent Endod. 2024 Jul;27(7):695-700. doi: 10.4103/JCDE.JCDE_189_24. Epub 2024 Jul 8. J Conserv Dent Endod. 2024. PMID: 39262599 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical