Continuous measurement of multiple inert and respiratory gas exchange in an anaesthetic breathing system by continuous indirect calorimetry
- PMID: 19169834
- DOI: 10.1007/s10877-009-9161-4
Continuous measurement of multiple inert and respiratory gas exchange in an anaesthetic breathing system by continuous indirect calorimetry
Abstract
A method was tested which permits continuous monitoring from a breathing system of the rate of uptake of multiple gas species, such as occurs in patients during inhalational anaesthesia. The method is an indirect calorimetry technique which uses fresh gas rotameters for control, regulation and measurement of the gas flows into the system, with continuous sampling of mixed exhaust gas, and frequent automated recalibration to maintain accuracy. Its accuracy was tested in 16 patients undergoing pre-cardiopulmonary bypass coronary artery surgery, breathing mixtures of oxygen/air and sevoflurane with/without nitrous oxide, by comparison with the reverse Fick method. Overall mean bias [95% confidence interval (CI)] of rate of uptake was 17.9 [7.3 to 28.5] ml min(-1) for oxygen, 0.04 [-0.42 to 0.50] ml min(-1) for sevoflurane, 10.9 [-16.1 to 37.8] for CO(2), and 8.8 [-14.8 to 32.4] ml min(-1) for nitrous oxide where present. The method proved to be accurate and precise, and allows continuous monitoring of exchange of multiple gases using standard gas analysis devices.
Similar articles
-
Non-invasive metabolic monitoring of patients under anaesthesia by continuous indirect calorimetry--an in vivo trial of a new method.Br J Anaesth. 2007 Jan;98(1):45-52. doi: 10.1093/bja/ael310. Epub 2006 Nov 22. Br J Anaesth. 2007. PMID: 17124187 Clinical Trial.
-
Physiologically precise simulation of multiple lung gas exchange during anaesthesia by simultaneous gas infusion and extraction.Physiol Meas. 2005 Dec;26(6):965-78. doi: 10.1088/0967-3334/26/6/007. Epub 2005 Oct 17. Physiol Meas. 2005. PMID: 16311445
-
Continuous indirect calorimetry--a laboratory simulation of a new method for non-invasive metabolic monitoring of patients under anaesthesia or in critical care.Physiol Meas. 2006 Feb;27(2):155-64. doi: 10.1088/0967-3334/27/2/006. Epub 2006 Jan 4. Physiol Meas. 2006. PMID: 16400202
-
[Nitrous oxide free low-flow anesthesia].Anaesthesiol Reanim. 2000;25(3):60-7. Anaesthesiol Reanim. 2000. PMID: 10920482 Review. German.
-
New and alternative delivery concepts and techniques.Best Pract Res Clin Anaesthesiol. 2005 Sep;19(3):415-28. doi: 10.1016/j.bpa.2005.02.002. Best Pract Res Clin Anaesthesiol. 2005. PMID: 16013691 Review.
Cited by
-
Measured Oxygen Consumption During Pediatric Cardiac Catheterization is More Accurate than Assumed Oxygen Consumption.Pediatr Cardiol. 2024 Oct;45(7):1466-1474. doi: 10.1007/s00246-023-03186-x. Epub 2023 May 27. Pediatr Cardiol. 2024. PMID: 37243747
-
Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures.BMC Res Notes. 2014 Jul 23;7:469. doi: 10.1186/1756-0500-7-469. BMC Res Notes. 2014. PMID: 25056252 Free PMC article. Clinical Trial.
-
In vitro performance of prefilled CO₂ absorbers with the Aisys®.J Clin Monit Comput. 2016 Apr;30(2):193-202. doi: 10.1007/s10877-015-9699-2. Epub 2015 May 8. J Clin Monit Comput. 2016. PMID: 25953416
References
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous