[Sensitivity and specificity of 4 domestic ELISA kits for detection of hepatitis B virus markers]

[Article in Chinese]
Affiliations
  • PMID: 19173858
Comparative Study

[Sensitivity and specificity of 4 domestic ELISA kits for detection of hepatitis B virus markers]

[Article in Chinese]
Xing Wu et al. Zhonghua Liu Xing Bing Xue Za Zhi. 2008 Sep.

Abstract

Objective: To compare and analyze the sensitivity, specificity of 4 domestic ELISA kits for detection of hepatitis B virus (HBV) markers (HBsAg, anti-HBs, HBeAg, anti-HBe, and anti-HBc).

Methods: Five hundred and ninety four serum samples collected from patients with chronic hepatitis B and abnormal blood donors were detected for HBV markers and by 4 domestic ELISA kits. Samples with conflicting results by different diagnostic kits were retested. Samples with the HBsAg values close to the cut-off point were detected by Abbott HBsAg confirmation kit (Architect HBsAg confirm). Sensitivity of the kits was determined, using the national sensitivity reference panels for HBsAg, anti-HBs, HBeAg, anti-HBe and anti-HBc.

Results: The rates of sensitivity on 4 domestic kits for detection of HBsAg were 4 to 10 times lower, and on the 4 domestic kits for detection of anti-HBs, HBeAg, anti-HBe and anti-HBc were 4 to 16 times lower, as compared to Abbott Architect kits. In addition, the domestic HBV ELISA kits had some false positive results. The total coincidence rates of HBsAg, anti-HBs, HBeAg, anti-HBe, anti-HBc were 96.46%-98.15%, 94.28%-98.15%, 98.15%-99.49%, 90.07%-96.30%, 92.09%-96.80%, respectively.

Conclusion: Both sensitivity and specificity of the domestically produced HBV ELISA kits should be improved.