Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy
- PMID: 1918102
- DOI: 10.1002/jbm.820250703
Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy
Abstract
Partially stabilized zirconia ceramic is being recognized among ceramics for its high strength and toughness. With this ceramic, is possible to manufacture a 22-mm-size femoral head for low friction arthroplasty of the hip joint in association with an ultra-high-molecular-weight polyethylene socket. Wear-resistant properties of zirconia ceramic were screened on two principally different wear devices. Sterile calf bovine serum, physiological saline, and distilled water were chosen as the lubricant fluid media. Depending on the lubricant medium, the wear factor of polyethylene against zirconia ceramic counterfaces was 40 to 60% less than that against alumina ceramic counterfaces, and 5 to 10 times lower than with the SUS316L metal counterfaces. Polyethylene wear against metal was more susceptible in saline in which it had 2 to 3 times higher wear rate than with serum. On the other hand, different fluid media had little effect on polyethylene wear against ceramic counterfaces. In each set of tests, the wear factor obtained on an unidirectional wear device showed 10 to 15 times higher values, in comparison to the wear factor estimated on a reciprocating wear device.
Similar articles
-
Friction and wear properties of polymer, metal, and ceramic prosthetic joint materials evaluated on a multichannel screening device.J Biomed Mater Res. 1981 Sep;15(5):619-53. doi: 10.1002/jbm.820150503. J Biomed Mater Res. 1981. PMID: 12659132
-
Wear properties of alumina/zirconia composite ceramics for joint prostheses measured with an end-face apparatus.Biomed Mater Eng. 2004;14(3):263-70. Biomed Mater Eng. 2004. PMID: 15299238
-
The wear of ultra-high molecular weight polyethylene sliding on metallic and ceramic counterfaces representative of current femoral surfaces in joint replacement.Proc Inst Mech Eng H. 1997;211(1):17-24. doi: 10.1243/0954411971534647. Proc Inst Mech Eng H. 1997. PMID: 9141887
-
Characteristics of metal and ceramic total hip bearing surfaces and their effect on long-term ultra high molecular weight polyethylene wear.Clin Orthop Relat Res. 1993 Sep;(294):361-78. Clin Orthop Relat Res. 1993. PMID: 8358943 Review.
-
Alternative materials to improve total hip replacement tribology.Acta Orthop Scand. 2003 Aug;74(4):380-8. doi: 10.1080/00016470310017668. Acta Orthop Scand. 2003. PMID: 14521286 Review.
Cited by
-
Lessons of a hip failure.BMJ. 1998 Feb 28;316(7132):644. doi: 10.1136/bmj.316.7132.644. BMJ. 1998. PMID: 9522775 Free PMC article. No abstract available.
-
Esthetic Prosthetic Restorations: Reliability and Effects on Antagonist Dentition.Open Dent J. 2015 Dec 31;9:473-81. doi: 10.2174/1874210601509010473. eCollection 2015. Open Dent J. 2015. PMID: 26962376 Free PMC article.
-
Corrosion behaviour and mechanical properties of functionally gradient materials developed for possible hard-tissue applications.J Mater Sci Mater Med. 1997 Dec;8(12):793-7. doi: 10.1023/a:1018525015421. J Mater Sci Mater Med. 1997. PMID: 15348793
-
Retrieval analysis of new-generation yttria-stabilized zirconia femoral heads after total hip arthroplasty.Eur J Orthop Surg Traumatol. 2014 Oct;24(7):1197-202. doi: 10.1007/s00590-013-1397-5. Epub 2013 Dec 31. Eur J Orthop Surg Traumatol. 2014. PMID: 24378655
-
Triborheological Study under Physiological Conditions of PVA Hydrogel/HA Lubricant as Synthetic System for Soft Tissue Replacement.Polymers (Basel). 2021 Feb 28;13(5):746. doi: 10.3390/polym13050746. Polymers (Basel). 2021. PMID: 33670837 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources