Artifacts in digital mammography
- PMID: 19203002
Artifacts in digital mammography
Abstract
In April 2005 screening with digital mammography was allowed in the Flemish part of Belgium. A rigorous physical-technical Quality Control (QC) procedure based on the European guidelines (EUREF) was then implemented. Besides quality control, there is also quality assurance (QA). Detection of artifacts is part of the QA. During the central second reading, a continuous evaluation of the image quality is done. All visible artifacts in the digital images are registered and collected. All systems participate also in a daily quality control, with a daily exposure of a phantom image which is sent to the certified quality control group. The collected artifacts were divided into 5 different categories: patient related artifacts, technologist related artifacts, mammography unit related artifacts, processing related artifacts and viewing conditions related artifacts. Patient related artifacts are comparable with film screen mammography (FSM) and are therefore not discussed. One of the main artifacts in the group of technologist related artifacts is dust in the cassette of computed radiography (CR) systems. In the group of mammography unit related artifacts a distinction is made between the artifacts of CR systems and direct radiography (DR) systems. In the CR group, the artifacts originate in the reader, whereas in the DR group they originate in the detector, which in our study was a Selenium detector. Artifacts due to failure of the Selenium detector are most frequent in this last group. Processing related artifacts are found when the reading of the processing algorithm by the system or by the PACS software made mistakes. Because there is a daily quality control of the monitors of the soft copy work stations, we didn't recognize viewing conditions related artifacts. Some of the artifacts can simulate breast lesions or can disturb the reading of the images. In order to avoid misinterpretation, recognizing artifacts and understanding their physical-technical background are of great importance in digital breast imaging.
Similar articles
-
Image quality assessment in digital mammography: part I. Technical characterization of the systems.Phys Med Biol. 2011 Jul 21;56(14):4201-20. doi: 10.1088/0031-9155/56/14/002. Epub 2011 Jun 23. Phys Med Biol. 2011. PMID: 21701051
-
An alternative method for noise analysis using pixel variance as part of quality control procedures on digital mammography systems.Phys Med Biol. 2009 Nov 21;54(22):6809-22. doi: 10.1088/0031-9155/54/22/004. Epub 2009 Oct 21. Phys Med Biol. 2009. PMID: 19847017
-
Soft-copy reading in digital mammography of mass: diagnostic performance of a 5-megapixel cathode ray tube monitor versus a 3-megapixel liquid crystal display monitor in a diagnostic setting.Acta Radiol. 2008 Jul;49(6):623-9. doi: 10.1080/02841850802022993. Acta Radiol. 2008. PMID: 18568553 Clinical Trial.
-
[Digital storage phosphor radiography. Doses and image quality].Radiol Med. 1994 Jun;87(6):847-51. Radiol Med. 1994. PMID: 8041940 Review. Italian.
-
Digital detectors in mammography. A technological overview.JBR-BTR. 2000 Apr;83(2):84-7. JBR-BTR. 2000. PMID: 10859905 Review.
Cited by
-
Mammographic artifacts on full-field digital mammography.J Digit Imaging. 2014 Apr;27(2):231-6. doi: 10.1007/s10278-013-9641-4. J Digit Imaging. 2014. PMID: 24113845 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical
Miscellaneous