Feasibility of simultaneous pre- and postfilter transcranial Doppler monitoring during carotid artery stenting
- PMID: 19216954
- DOI: 10.1016/j.jvs.2008.08.102
Feasibility of simultaneous pre- and postfilter transcranial Doppler monitoring during carotid artery stenting
Abstract
Objective: Carotid artery stenting (CAS) is emerging as an acceptable treatment alternative to surgery for patients with carotid artery stenosis. The major risk of CAS is cerebral embolization of plaque and thrombus causing stroke or asymptomatic brain infarction. Use of embolic protection devices (EPD) to trap emboli before they reach the brain is now standard practice in CAS. The pore size of the currently available filters is >100 microns and emboli smaller than the EPD pores can still reach the brain. While the use of EPD is widespread, little evidence exists of their in vivo efficacy in preventing distal embolization. Our aim was to quantify the number of emboli reaching the brain with the device in place. Therefore, the expected value of this report is in its description of a novel application of transcranial Doppler (TCD). Due to the limited number of cases, it is not intended to support the use of one EPD over another.
Methods: Six patients were monitored with ipsilateral simultaneous dual probe TCD during CAS. Two types of cerebral protection systems were evaluated: FilterWire EZ System (FW; Boston Scientific, Santa Clara, Calif) and GORE Neuro Protection System (NPS; W.L. Gore and Associates, Flagstaff, Ariz). By placing TCD probes both proximal and distal to the filterwire EPD, we quantified the microembolic signals before the EPD as well as those, which reached the intracranial circulation after the EPD. One probe was placed submandibularly to monitor the ICA (SICA), while another was placed transtemporally to monitor the middle and anterior cerebral artery (MCA + ACA). We compare the number of extracranial emboli prior to the EPD with the number of intracranial emboli after the EPD.
Results: Dual probe monitoring was successful during the five stages of the CAS: lesion crossing (LC), predilatation (PreD), stent placement (SP), postdilatation (PostD), and filter/device removal (FR/DR). Using FW during LC by probe 1 (SICA)/probe 2 (MCA + ACA): (18 [range, 15-22]/15 [range, 11-20]), PreD (111 [range, 101-121]/101 [range, 90-111]), SP (68 [range, 60-76]/42 [range, 30-53]), PostD (27 [range, 25-30]/24 [range, 22-27]), FR (0.3 [range, 0-1]/0.7 [range, 0-1]) average number of microembolic signals were detected. Using NPS during LC (1.7 [range, 0-3]/1 [range, 0-2]), PreD (0/1.7 [range, 0-4]), SP (0/0), PostD (0/0), DR (18 [range, 0-18]/6.7 [range, 1-13]) average number of microembolic signals were detected.
Conclusion: EPD significantly reduces but does not eliminate the number of microemboli reaching the brain during carotid artery angioplasty and stenting. We propose monitoring of CAS with submandibular and transtemporal TCD probes to further evaluate the practice of distal embolization protection. Although our study is not powered to make any recommendations about EPDs, we believe that sequential dual probe TCD monitoring is a worthy tool with the potential to give vital information to assess the various devices and the techniques of utilization.
Similar articles
-
Atherosclerotic plaque composition assessed by virtual histology intravascular ultrasound and cerebral embolization after carotid stenting.J Vasc Surg. 2010 Nov;52(5):1188-94. doi: 10.1016/j.jvs.2010.05.101. Epub 2010 Jul 23. J Vasc Surg. 2010. PMID: 20655687
-
Embolism to the brain during carotid stenting and surgery.Acta Chir Belg. 2007 Mar-Apr;107(2):151-4. Acta Chir Belg. 2007. PMID: 17515263
-
Transcranial Doppler monitoring during stenting of the carotid bifurcation: evaluation of two different distal protection devices in preventing embolization.J Endovasc Ther. 2006 Aug;13(4):436-42. doi: 10.1583/05-1804MR.1. J Endovasc Ther. 2006. PMID: 16928156 Clinical Trial.
-
Evidence overview: benefit of cerebral protection devices during carotid artery stenting.J Cardiovasc Surg (Torino). 2017 Apr;58(2):170-177. doi: 10.23736/S0021-9509.16.09848-7. Epub 2016 Dec 22. J Cardiovasc Surg (Torino). 2017. PMID: 28004899 Review.
-
Filter devices for cerebral protection during carotid angioplasty and stenting.J Endovasc Ther. 2003 Dec;10(6):1039-45. doi: 10.1177/152660280301000603. J Endovasc Ther. 2003. PMID: 14723572 Review.
Cited by
-
Risk factors predictive of carotid artery stenting-associated subclinical microemboli.Int J Angiol. 2011 Mar;20(1):25-32. doi: 10.1055/s-0031-1272546. Int J Angiol. 2011. PMID: 22532767 Free PMC article.
-
Microembolization is associated with transient cognitive decline in patients undergoing carotid interventions.J Vasc Surg. 2016 Dec;64(6):1719-1725. doi: 10.1016/j.jvs.2016.06.104. Epub 2016 Sep 12. J Vasc Surg. 2016. PMID: 27633169 Free PMC article.
-
Safety and Efficacy of Flow Reversal in Acute and Elective Carotid Angioplasty and Stenting Using the Mo.Ma Device with Short-Term Follow-Up.Interv Neurol. 2020 Jan;8(2-6):196-205. doi: 10.1159/000499045. Epub 2019 Aug 5. Interv Neurol. 2020. PMID: 32508902 Free PMC article.
-
Volume of subclinical embolic infarct correlates to long-term cognitive changes after carotid revascularization.J Vasc Surg. 2017 Mar;65(3):686-694. doi: 10.1016/j.jvs.2016.09.057. Epub 2016 Dec 23. J Vasc Surg. 2017. PMID: 28024850 Free PMC article.
-
Symptomatic carotid vasospasm caused by a distal-protection device during stent angioplasty of the right internal carotid artery.Tex Heart Inst J. 2010;37(2):226-9. Tex Heart Inst J. 2010. PMID: 20401301 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous