Marginal bone level changes and prosthetic maintenance of mandibular overdentures supported by 2 implants: a 5-year randomized clinical trial
- PMID: 19220845
- DOI: 10.1111/j.1708-8208.2008.00143.x
Marginal bone level changes and prosthetic maintenance of mandibular overdentures supported by 2 implants: a 5-year randomized clinical trial
Abstract
Background: Documentation of early loading of mandibular overdentures supported by different implant systems is scarce.
Purpose: This study aimed to compare the biologic and prosthetic outcome of mandibular overdentures supported by unsplinted early-loaded one- and two-stage oral implants after 5 years of function.
Materials and methods: Twenty-eight consecutive patients were screened following an inclusion and exclusion criteria, and randomly allocated to treatment groups. Ball-retained mandibular overdentures were fabricated on two unsplinted Straumann (Institut Straumann AG, Basel, Switzerland) and Brånemark (Nobel Biocare AB, Göteborg, Sweden) dental implants and subjected to an early-loading protocol. During the 5-year period, prosthetic complications were recorded. At 5-years of function, plaque, peri-implant inflammation, bleeding, and calculus index scores were recorded, and standard periapical radiographs were obtained from each implant for measurement of marginal bone loss.
Results: All implants survived during the observation period. The peri-implant inflammation, bleeding, and calculus index scores around Straumann and Brånemark implants were similar (p > .05). The marginal bone loss around Brånemark implants (1.21 +/- 0.1) was higher than Straumann implants (0.73 +/- 0.06) at 5 years of function (p = .002). Kaplan-Meier tests revealed that 1- and 5-year survival of overdentures on Straumann and Brånemark implants were similar (p = .85). Wear of the ball abutment in the Brånemark group was higher than in the Straumann group (p < .05). Complications regarding the retainer and the need for occlusal adjustments were higher in the Straumann group (p < .05). Chi-square test revealed that the frequency of retightening of the retainer was higher in the Straumann group than in the Brånemark group (p < .05).
Conclusions: Mandibular overdentures supported by unsplinted early-loaded Straumann and Brånemark implants lead to similar peri-implant soft tissue and prosthetic outcomes, although higher marginal bone loss could be observed around Brånemark implants after 5 years.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
