Delayed diagnosis of low-symptomatic ceramic acetabular liner fracture in ceramic-on-ceramic total hip arthroplasty
- PMID: 19226001
Delayed diagnosis of low-symptomatic ceramic acetabular liner fracture in ceramic-on-ceramic total hip arthroplasty
Abstract
Alumina ceramics in total hip arthroplasty (THA) have been used in Europe since 1970. Over the years, ceramic-on-ceramic coupling became a valid option in THA because of excellent biocompatibility and tribological properties. The major disadvantages are possible squeaking and risk of breakage, usually disclosed by pain and functional impairment. Squeaking is an audible noise arising from ceramic-on-ceramic bearings, the incidence of which is reported to range from 1% to 7% of THAs. Component positioning, stripe wear, and edge loading have all been implicated. Clicking sounds and scratching have also been anecdotally described. Breakage of a ceramic component due to brittleness of the material is a rare complication. Trauma, high activity level, and obesity may increase the risk of fracture. Defective ceramic manufacture, inadequate implant design, and errors in surgical technique may contribute to breakage. Asian population lifestyle, including squatting, kneeling, and sitting cross-legged, has been correlated to liner rim impingement and fracture. Additional reports concerning failure of various ceramic liners have recently been described.
Similar articles
-
Fracture of a ceramic liner in a total hip arthroplasty with a sandwich cup.Arch Orthop Trauma Surg. 2008 Aug;128(8):783-5. doi: 10.1007/s00402-007-0536-z. Epub 2007 Dec 8. Arch Orthop Trauma Surg. 2008. PMID: 18066561
-
Alumina femoral head fracture in uncemented total hip arthroplasty with a ceramic sandwich cup.J Arthroplasty. 2007 Apr;22(3):468-71. doi: 10.1016/j.arth.2006.05.020. J Arthroplasty. 2007. PMID: 17400106
-
Fracture of a titanium sleeve-encased third-generation ceramic liner in a modern THA.Orthopedics. 2011 Oct 5;34(10):e682-4. doi: 10.3928/01477447-20110826-30. Orthopedics. 2011. PMID: 21956067
-
Alternative materials to improve total hip replacement tribology.Acta Orthop Scand. 2003 Aug;74(4):380-8. doi: 10.1080/00016470310017668. Acta Orthop Scand. 2003. PMID: 14521286 Review.
-
Advantages and disadvantages of ceramic on ceramic total hip arthroplasty: a review.Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2012 Sep;156(3):204-12. doi: 10.5507/bp.2012.063. Epub 2012 Oct 4. Biomed Pap Med Fac Univ Palacky Olomouc Czech Repub. 2012. PMID: 23069885 Review.
Cited by
-
Noise production in ceramic-on-ceramic total hip arthroplasty is associated with lower patient satisfaction and hip scores.J Orthop. 2016 Jun 29;13(4):282-284. doi: 10.1016/j.jor.2016.06.018. eCollection 2016 Dec. J Orthop. 2016. PMID: 27408504 Free PMC article.
-
Catastrophic failure of ceramic-on-ceramic total hip arthroplasty presenting as squeaking hip.BMJ Case Rep. 2013 Feb 20;2013:bcr2013008614. doi: 10.1136/bcr-2013-008614. BMJ Case Rep. 2013. PMID: 23429031 Free PMC article.
-
Effect of Alumina Particles on the Osteogenic Ability of Osteoblasts.J Funct Biomater. 2022 Jul 28;13(3):105. doi: 10.3390/jfb13030105. J Funct Biomater. 2022. PMID: 35997443 Free PMC article.
-
Catastrophic ceramic failure in total hip arthroplasty: the role of microseparation.JRSM Short Rep. 2011 Dec;2(12):96. doi: 10.1258/shorts.2011.011108. Epub 2011 Dec 15. JRSM Short Rep. 2011. PMID: 22279606 Free PMC article. No abstract available.
-
Nine years follow-up of 100 ceramic-on-ceramic total hip arthroplasty.Int Orthop. 2011 Nov;35(11):1599-604. doi: 10.1007/s00264-010-1185-3. Epub 2010 Dec 21. Int Orthop. 2011. PMID: 21174208 Free PMC article.