Evaluation of flow characteristics of 3-way catheters
- PMID: 19237172
- DOI: 10.1016/j.juro.2008.11.094
Evaluation of flow characteristics of 3-way catheters
Abstract
Purpose: We evaluated the flow characteristics of various 3-way catheters.
Materials and methods: We evaluated 14, 3-way catheters, including 18Fr, 20Fr, 22Fr and 24Fr Rusch(R), 18Fr, 20Fr, 22Fr and 24Fr Bardex(R), 18Fr, 20Fr, 22Fr and 24Fr Dover, and 20Fr and 22Fr Mentor (Mentor, Santa Barbara, California) catheters. Ten operators applied maximum 1-hand pressure with a 60 ml bladder wash syringe. Maximum and average flow rates were evaluated. The catheter was connected to an irrigation bottle and free flow irrigation speed was noted. The catheter was inserted into a latex balloon containing a prefilled volume of 100 ml and continuous irrigation characteristics were noted. Statistical analysis was performed using 1-way ANOVA and the Kruskal-Wallis test.
Results: Rusch catheters had the best flow characteristics when the drainage port was used for manual irrigation with 18Fr catheters (Rusch and Dover 24.8 and 19.2 ml per second, p = 0.003, vs Bardex 17.67 ml per second, p <0.001), and with 20Fr Rusch, Bardex, Mentor and Dover catheters (27.7, 21.42, 27.1 and 24 ml per second, respectively, p = 0.034). In the other categories of catheters tested there was no significant difference among 22Fr Rusch, Bardex, Dover and Mentor catheters (29.4, 28.9, 25 and 28.27 ml per second, p = 0.32), and among 24Fr Rusch, Bardex and Dover catheters (32.2, 29.79 and 29.9 ml per second, respectively, p = 0.27). Upon using the irrigation channel for manual irrigation all catheters had similar flow characteristics (no statistically significant difference). When connected to the irrigation tube with free flow, although the 18Fr, 20Fr and 22Fr Rusch, and 24Fr Dover catheters had slightly better flow than the others, this was not statistically significant. There was no marked difference in flow rate as catheter size increased above 20Fr. When the artificial bladder was used, the Rusch catheters had the maximum drainage in the 18Fr and 20Fr sizes, whereas the Mentor and Dover catheters had the maximum drainage in the 22Fr and 24Fr sizes, respectively (no statistically significant difference).
Conclusions: The 18Fr and 20Fr Rusch 3-way catheters have better flow than other catheters when the drainage port is used for washout. In the 22Fr and 24Fr categories all different catheters had equivalent irrigation and drainage properties. Larger catheter size does not equate to better irrigation or drainage when continuous irrigation is used.
Similar articles
-
Irrigation and drainage properties of three-way urethral catheters.Urology. 2006 Jan;67(1):40-4. doi: 10.1016/j.urology.2005.07.007. Urology. 2006. PMID: 16413329
-
Analysis of flow rate of continuous bladder irrigation according to the height of the irrigation infusion set.Sci Rep. 2023 Nov 12;13(1):19715. doi: 10.1038/s41598-023-47198-2. Sci Rep. 2023. PMID: 37953276 Free PMC article.
-
The Variable Flow Characteristics for Different Brands of 3-Way Urinary Catheters: Proposing an Alternate and Accurate Standardised Labelling System.Urology. 2016 Mar;89:155-9. doi: 10.1016/j.urology.2015.12.022. Epub 2016 Jan 6. Urology. 2016. PMID: 26768713
-
A new model for suprapubic catheterization: the MediPlus Seldinger suprapubic catheter.Expert Rev Med Devices. 2008 Nov;5(6):705-7. doi: 10.1586/17434440.5.6.705. Expert Rev Med Devices. 2008. PMID: 19025347 Review.
-
Management of intra-abdominal abscess due to surgical site infection.Surg Technol Int. 2006;15:37-9. Surg Technol Int. 2006. PMID: 17029159 Review.
Cited by
-
Evaluation of urinary catheters for effective manual bladder washout.Sci Rep. 2022 Aug 23;12(1):14359. doi: 10.1038/s41598-022-18778-5. Sci Rep. 2022. PMID: 35999265 Free PMC article.
-
A comparative assessment of irrigation and drainage characteristics for commercially available urethral catheters.Cent European J Urol. 2017;70(4):382-387. doi: 10.5173/ceju.2017.1494. Epub 2017 Oct 3. Cent European J Urol. 2017. PMID: 29410890 Free PMC article.
-
Same sized three-way indwelling urinary catheters from various manufacturers present different irrigation and drainage properties.Ther Adv Urol. 2020 Jan 9;12:1756287219889496. doi: 10.1177/1756287219889496. eCollection 2020 Jan-Dec. Ther Adv Urol. 2020. PMID: 31949476 Free PMC article.
-
An improved delivery system for bladder irrigation.Ther Clin Risk Manag. 2010 Oct 5;6:459-62. doi: 10.2147/TCRM.S13525. Ther Clin Risk Manag. 2010. PMID: 20957138 Free PMC article.