A clinical comparison of the Oculab Tono-Pen with the Goldmann applanation tonometer in eyes filled with silicone oil
- PMID: 1925086
- DOI: 10.1097/00006982-199111020-00005
A clinical comparison of the Oculab Tono-Pen with the Goldmann applanation tonometer in eyes filled with silicone oil
Abstract
Intraocular pressure in 21 eyes filled with silicone oil was measured with two different instruments: the Oculab Tono-Pen and a Goldmann applanation tonometer mounted to a slit lamp biomicroscope. Intraocular pressure ranged from 2 mmHg to 28 mmHg, and the mean difference between readings obtained from the two instruments was 0.64, which was not statistically significant. These results suggest that the Oculab Tono-Pen is an effective instrument with which to measure intraocular pressure in silicone-filled eye.