Range of motion in total knee arthroplasty: a prospective comparison of high-flexion and standard cruciate-retaining designs
- PMID: 19255229
- DOI: 10.2106/JBJS.H.00300
Range of motion in total knee arthroplasty: a prospective comparison of high-flexion and standard cruciate-retaining designs
Abstract
Background: Range of motion after a total knee arthroplasty is an important indicator of clinical outcome. Recently, a high-flexion posterior cruciate ligament-retaining knee prosthesis was designed to allow greater flexion after total knee arthroplasty. The purpose of this study was to compare range of motion and functional outcomes in patients who received either a high-flexion cruciate-retaining or a standard cruciate-retaining knee replacement.
Methods: Fifty knees that had a total knee arthroplasty with a high-flexion design and fifty that had a total knee arthroplasty with a standard design were included in this study and were followed prospectively for a minimum of two years. The arcs of maximal non-weight-bearing passive flexion and weight-bearing flexion were measured, and the number of knees that allowed the patients to kneel and sit cross-legged in comfort was determined. In addition, the functional outcomes in these two groups were assessed with use of the Hospital for Special Surgery and Western Ontario and McMaster Universities Osteoarthritis Index scores.
Results: At the time of the final follow-up, the average maximal non-weight-bearing flexion was 135.3 degrees for the knees in the high-flexion group and 134.3 degrees for the knees in the standard group; the difference was not significant. Moreover, no significant difference was found between the groups in terms of weight-bearing flexion (124.8 degrees in the high-flexion group and 123.7 degrees in the standard group) and the number of knees that allowed kneeling and sitting cross-legged. The average Hospital for Special Surgery knee score was 94.4 points in the high-flexion group and 92.4 points in the standard group; the difference was not significant. The Western Ontario and McMaster Universities Osteoarthritis Index scores also showed no significant difference between the groups.
Conclusions: For knees managed with a cruciate-retaining total knee arthroplasty, those that had the high-flexion design and those that had the standard design were found to have a similar range of motion under both non-weight-bearing and weight-bearing conditions. Moreover, no significant difference was found in terms of the other functional outcomes examined.
Similar articles
-
Functional outcome and range of motion of high-flexion posterior cruciate-retaining and high-flexion posterior cruciate-substituting total knee prostheses. A prospective, randomized study.J Bone Joint Surg Am. 2009 Apr;91(4):753-60. doi: 10.2106/JBJS.H.00805. J Bone Joint Surg Am. 2009. PMID: 19339558 Clinical Trial.
-
High-flexion posterior-substituting versus cruciate-retaining prosthesis in total knee arthroplasty: functional outcome, range of motion and complication comparison.Arch Orthop Trauma Surg. 2015 Jan;135(1):119-24. doi: 10.1007/s00402-014-2107-4. Epub 2014 Nov 12. Arch Orthop Trauma Surg. 2015. PMID: 25388862
-
Range of motion of standard and high-flexion posterior cruciate-retaining total knee prostheses a prospective randomized study.J Bone Joint Surg Am. 2009 Aug;91(8):1874-81. doi: 10.2106/JBJS.H.00769. J Bone Joint Surg Am. 2009. PMID: 19651944 Clinical Trial.
-
Influence of component design on in vivo tibiofemoral contact patterns during kneeling after total knee arthroplasty: a systematic review and meta-analysis.Knee Surg Sports Traumatol Arthrosc. 2021 Feb;29(2):446-466. doi: 10.1007/s00167-020-05949-y. Epub 2020 Apr 3. Knee Surg Sports Traumatol Arthrosc. 2021. PMID: 32242268
-
Review article: knee flexion after total knee arthroplasty.J Orthop Surg (Hong Kong). 2002 Dec;10(2):194-202. doi: 10.1177/230949900201000215. J Orthop Surg (Hong Kong). 2002. PMID: 12493934 Review.
Cited by
-
High flexion total knee arthroplasty - mid-term follow up of 5 years.Open Orthop J. 2011 Apr 14;5:138-42. doi: 10.2174/1874325001105010138. Open Orthop J. 2011. PMID: 21584199 Free PMC article.
-
High-flexion total knee arthroplasty improves flexion of stiff knees.Knee Surg Sports Traumatol Arthrosc. 2011 Jun;19(6):936-42. doi: 10.1007/s00167-010-1272-4. Epub 2010 Oct 2. Knee Surg Sports Traumatol Arthrosc. 2011. PMID: 20890698
-
Clinical, Functional, and Midterm Survival Analysis on Sigma Curved Plus Ultracongruent Polyethylene Insert in Primary Total Knee Arthroplasty: A Retrospective Study.Cureus. 2020 Nov 17;12(11):e11519. doi: 10.7759/cureus.11519. Cureus. 2020. PMID: 33354463 Free PMC article.
-
Comparison of anterior and rotatory laxity using navigation between single- and double-bundle ACL reconstruction: prospective randomized trial.Knee Surg Sports Traumatol Arthrosc. 2012 Apr;20(4):752-61. doi: 10.1007/s00167-012-1880-2. Epub 2012 Jan 25. Knee Surg Sports Traumatol Arthrosc. 2012. PMID: 22273781 Clinical Trial.
-
Cultural and urban city living expectations of knee motion in a Southeast Asian city: implications on total knee arthroplasty outcomes.Front Rehabil Sci. 2025 Jan 7;5:1446389. doi: 10.3389/fresc.2024.1446389. eCollection 2024. Front Rehabil Sci. 2025. PMID: 39844901 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources