Abstract
Objective: To present a novel algorithm for estimating recruitable alveolar collapse and hyperdistension based on electrical impedance tomography (EIT) during a decremental positive end-expiratory pressure (PEEP) titration.
Design: Technical note with illustrative case reports.
Setting: Respiratory intensive care unit.
Patient: Patients with acute respiratory distress syndrome.
Interventions: Lung recruitment and PEEP titration maneuver.
Measurements and results: Simultaneous acquisition of EIT and X-ray computerized tomography (CT) data. We found good agreement (in terms of amount and spatial location) between the collapse estimated by EIT and CT for all levels of PEEP. The optimal PEEP values detected by EIT for patients 1 and 2 (keeping lung collapse <10%) were 19 and 17 cmH(2)O, respectively. Although pointing to the same non-dependent lung regions, EIT estimates of hyperdistension represent the functional deterioration of lung units, instead of their anatomical changes, and could not be compared directly with static CT estimates for hyperinflation.
Conclusions: We described an EIT-based method for estimating recruitable alveolar collapse at the bedside, pointing out its regional distribution. Additionally, we proposed a measure of lung hyperdistension based on regional lung mechanics.
Similar articles
-
Electrical impedance tomography: A compass for the safe route to optimal PEEP.Respir Med. 2021 Oct;187:106555. doi: 10.1016/j.rmed.2021.106555. Epub 2021 Jul 30. Respir Med. 2021. PMID: 34352563 Review.
-
Assessment of electrical impedance tomography to set optimal positive end-expiratory pressure for veno-venous ECMO-treated severe ARDS patients.J Crit Care. 2020 Dec;60:38-44. doi: 10.1016/j.jcrc.2020.06.017. Epub 2020 Jul 8. J Crit Care. 2020. PMID: 32736198
-
Variation of poorly ventilated lung units (silent spaces) measured by electrical impedance tomography to dynamically assess recruitment.Crit Care. 2018 Jan 31;22(1):26. doi: 10.1186/s13054-017-1931-7. Crit Care. 2018. PMID: 29386048 Free PMC article. Clinical Trial.
-
A novel positive end-expiratory pressure titration using electrical impedance tomography in spontaneously breathing acute respiratory distress syndrome patients on mechanical ventilation: an observational study from the MaastrICCht cohort.J Clin Monit Comput. 2025 Feb;39(1):127-139. doi: 10.1007/s10877-024-01212-8. Epub 2024 Aug 28. J Clin Monit Comput. 2025. PMID: 39196479 Free PMC article.
-
How to assess positive end-expiratory pressure-induced alveolar recruitment?Minerva Anestesiol. 2013 Jan;79(1):83-91. Epub 2012 Nov 8. Minerva Anestesiol. 2013. PMID: 23135694 Review.
Cited by
-
Heart-lung crosstalk in acute respiratory distress syndrome.Front Physiol. 2024 Oct 18;15:1478514. doi: 10.3389/fphys.2024.1478514. eCollection 2024. Front Physiol. 2024. PMID: 39493867 Free PMC article. Review.
-
Lung aeration changes after lung recruitment in children with acute lung injury: a feasibility study.Pediatr Pulmonol. 2012 Aug;47(8):771-9. doi: 10.1002/ppul.22508. Epub 2012 Feb 1. Pediatr Pulmonol. 2012. PMID: 22298419 Free PMC article.
-
Prospective Observational Study to Evaluate the Effect of Different Levels of Positive End-Expiratory Pressure on Lung Mechanics in Patients with and without Acute Respiratory Distress Syndrome.J Clin Med. 2020 Jul 31;9(8):2446. doi: 10.3390/jcm9082446. J Clin Med. 2020. PMID: 32751791 Free PMC article.
-
Clinical Scenarios of the Application of Electrical Impedance Tomography in Paediatric Intensive Care.Sci Rep. 2019 Mar 29;9(1):5362. doi: 10.1038/s41598-019-41774-1. Sci Rep. 2019. PMID: 30926828 Free PMC article.
-
Clinical Applicability of Electrical Impedance Tomography in Patient-Tailored Ventilation: A Narrative Review.Tomography. 2023 Oct 18;9(5):1903-1932. doi: 10.3390/tomography9050150. Tomography. 2023. PMID: 37888742 Free PMC article. Review.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical