Computed tomography during cardiopulmonary resuscitation using automated chest compression devices--an initial study
- PMID: 19259679
- DOI: 10.1007/s00330-009-1359-9
Computed tomography during cardiopulmonary resuscitation using automated chest compression devices--an initial study
Abstract
The purpose of the study was to evaluate both CT image quality in a phantom study and feasibility in an initial case series using automated chest compression (A-CC) devices for cardiopulmonary resuscitation (CPR). Multidetector CT (MDCT) of a chest/heart phantom (Thorax-CCI, QRM, Germany) was performed with identical protocols of the phantom alone (S), the phantom together with two different A-CC devices (A: AutoPulse, Zoll, Germany; L: LUCAS, Jolife, Sweden), and the phantom with a LUCAS baseplate, but without the compression unit (L-bp). Nine radiologists evaluated image noise quantitatively (n = 244 regions, Student's t-test) and also rated image quality subjectively (1-excellent to 6-inadequate, Mann-Whitney U-test). Additionally, three patients during prolonged CPR underwent CT with A-CC devices. Mean image noise of S was increased by 1.21 using L-bp, by 3.62 using A, and by 5.94 using L (p < 0.01 each). Image quality was identical using S and L-bp (1.64 each), slightly worse with A (1.83), and significantly worse with L (2.97, p < 0.001). In all patient cases the main lesions were identified, which led to clinical key decisions. Image quality was excellent with L-bp and good with A. Under CPR conditions initial cases indicate that MDCT diagnostics supports either focused treatment or the decision to terminate efforts.
Similar articles
-
Computed Tomography Imaging in Simulated Ongoing Cardiopulmonary Resuscitation: No Need to Switch Off the Chest Compression Device during Image Acquisition.Diagnostics (Basel). 2021 Jun 19;11(6):1122. doi: 10.3390/diagnostics11061122. Diagnostics (Basel). 2021. PMID: 34205468 Free PMC article.
-
Multidetector CT and radiographic findings of lung injuries secondary to cardiopulmonary resuscitation.Injury. 2013 Sep;44(9):1204-7. doi: 10.1016/j.injury.2012.10.004. Epub 2013 Jan 29. Injury. 2013. PMID: 23374163
-
Evaluating the Quality of Cardiopulmonary Resuscitation in the Emergency Department by Real-Time Video Recording System.PLoS One. 2015 Oct 2;10(10):e0139825. doi: 10.1371/journal.pone.0139825. eCollection 2015. PLoS One. 2015. PMID: 26431420 Free PMC article.
-
[The use of mechanical chest compression devices for both out-of-hospital and in-hospital refractory cardiac arrest].G Ital Cardiol (Rome). 2017 Apr;18(4):305-312. doi: 10.1714/2683.27473. G Ital Cardiol (Rome). 2017. PMID: 28492570 Review. Italian.
-
Cardiac Resuscitation RESEARCH REVIEW. Studies address the efficacy of continuous chest conpressions and automated cornmpression devices.EMS World. 2016 Feb;45(2):27-9. EMS World. 2016. PMID: 26946587 Review. No abstract available.
Cited by
-
Computed Tomography Imaging in Simulated Ongoing Cardiopulmonary Resuscitation: No Need to Switch Off the Chest Compression Device during Image Acquisition.Diagnostics (Basel). 2021 Jun 19;11(6):1122. doi: 10.3390/diagnostics11061122. Diagnostics (Basel). 2021. PMID: 34205468 Free PMC article.
-
Part 1: Executive summary: 2010 International Consensus on Cardiopulmonary Resuscitation and Emergency Cardiovascular Care Science With Treatment Recommendations.Resuscitation. 2010 Oct;81 Suppl 1(1):e1-25. doi: 10.1016/j.resuscitation.2010.08.002. Resuscitation. 2010. PMID: 20956042 Free PMC article. No abstract available.
-
Mechanical CPR: Who? When? How?Crit Care. 2018 May 29;22(1):140. doi: 10.1186/s13054-018-2059-0. Crit Care. 2018. PMID: 29843753 Free PMC article. Review.
-
Automated Chest Compression Allows Sufficient Resuscitation During Computed Tomography to Reveal the Cause of In-hospital Cardiac Arrest-A Case Report.Turk J Anaesthesiol Reanim. 2021 Jun;49(3):254-256. doi: 10.5152/TJAR.2021.318. Turk J Anaesthesiol Reanim. 2021. PMID: 35110147 Free PMC article.
-
[Importance of multidetector CT imaging in multiple trauma].Radiologe. 2014 Sep;54(9):861-71. doi: 10.1007/s00117-013-2634-y. Radiologe. 2014. PMID: 25216568 German.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous