Case Reports
. 2009 Aug;19(8):1857-66.
doi: 10.1007/s00330-009-1359-9. Epub 2009 Mar 4.

Computed tomography during cardiopulmonary resuscitation using automated chest compression devices--an initial study

Affiliations
Case Reports

Computed tomography during cardiopulmonary resuscitation using automated chest compression devices--an initial study

Stefan Wirth et al. Eur Radiol. 2009 Aug.

Abstract

The purpose of the study was to evaluate both CT image quality in a phantom study and feasibility in an initial case series using automated chest compression (A-CC) devices for cardiopulmonary resuscitation (CPR). Multidetector CT (MDCT) of a chest/heart phantom (Thorax-CCI, QRM, Germany) was performed with identical protocols of the phantom alone (S), the phantom together with two different A-CC devices (A: AutoPulse, Zoll, Germany; L: LUCAS, Jolife, Sweden), and the phantom with a LUCAS baseplate, but without the compression unit (L-bp). Nine radiologists evaluated image noise quantitatively (n = 244 regions, Student's t-test) and also rated image quality subjectively (1-excellent to 6-inadequate, Mann-Whitney U-test). Additionally, three patients during prolonged CPR underwent CT with A-CC devices. Mean image noise of S was increased by 1.21 using L-bp, by 3.62 using A, and by 5.94 using L (p < 0.01 each). Image quality was identical using S and L-bp (1.64 each), slightly worse with A (1.83), and significantly worse with L (2.97, p < 0.001). In all patient cases the main lesions were identified, which led to clinical key decisions. Image quality was excellent with L-bp and good with A. Under CPR conditions initial cases indicate that MDCT diagnostics supports either focused treatment or the decision to terminate efforts.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Resuscitation. 2005 Jun;65(3):357-63 - PubMed
    1. Unfallchirurg. 2004 Oct;107(10):937-44 - PubMed
    1. Resuscitation. 2004 Jun;61(3):273-80 - PubMed
    1. J Am Coll Cardiol. 2004 Dec 7;44(11):2214-20 - PubMed
    1. JAMA. 2006 Jun 14;295(22):2629-37 - PubMed

Publication types

MeSH terms