The influence of liquid crystal display monitors on observer performance for the detection of interstitial lung markings on both storage phosphor and flat-panel-detector chest radiography
- PMID: 19304429
- DOI: 10.1016/j.ejrad.2009.02.009
The influence of liquid crystal display monitors on observer performance for the detection of interstitial lung markings on both storage phosphor and flat-panel-detector chest radiography
Abstract
Purpose: To compare observer performance with a flat-panel liquid crystal display (LCD) monitor and with a high-resolution gray-scale cathode ray tube (CRT) monitor in the detection of interstitial lung markings using a silicon flat-panel-detector direct radiography (DR) and storage phosphor computed radiography (CR) in a clinical setting.
Materials and methods: We displayed 39 sets of posteroanterior chest radiographs from the patients who were suspected of interstitial lung disease. Each sets consisted of DR, CR and thin-section CT as the reference standard. Image identities were masked, randomly sorted, and displayed on both five mega pixel (2048x2560x8 bits) LCD and CRT monitors. Ten radiologists independently rated their confidence in detection for the presence of linear opacities in the four fields of the lungs; right upper, left upper, right lower, and left lower quadrant. Performance of a total 6240 (39 setsx2 detector systemsx2 monitor systemx4 fieldsx10 observers) observations was analyzed by multi-reader multi-case receiver operating characteristic (ROC) analysis. Differences between monitor systems in combinations of detector systems were compared using ANOVA and paired-samples t-test.
Results: Area under curves (AUC) for the presence of linear opacities measured by ROC analysis was higher on the LCDs than CRTs without statistical significance (p=0.082). AUC was significantly higher on the DR systems than CR systems (p=0.006). AUC was significantly higher on the LCDs than CRTs for DR systems (p=0.039) but not different for CR systems (p=0.301).
Conclusion: In clinical conditions, performance of the LCD monitor appears to be better for detecting interstitial lung markings when interfaced with DR systems.
Copyright (c) 2010 Elsevier Ireland Ltd. All rights reserved.
Similar articles
-
[Comparison of LCD and CRT monitors for detection of pulmonary nodules and interstitial lung diseases on digital chest radiographs by using receiver operating characteristic analysis].Nihon Hoshasen Gijutsu Gakkai Zasshi. 2006 May 20;62(5):734-41. doi: 10.6009/jjrt.62.734. Nihon Hoshasen Gijutsu Gakkai Zasshi. 2006. PMID: 16733502 Japanese.
-
Comparison of observer performance on soft-copy reading of digital chest radiographs: high resolution liquid-crystal display monitors versus cathode-ray tube monitors.Eur J Radiol. 2008 Apr;66(1):13-8. doi: 10.1016/j.ejrad.2007.05.023. Epub 2007 Aug 6. Eur J Radiol. 2008. PMID: 17689217
-
Detectability of catheters on bedside chest radiographs: comparison between liquid crystal display and high-resolution cathode-ray tube monitors.Radiology. 2005 Feb;234(2):611-6. doi: 10.1148/radiol.2342031297. Epub 2004 Dec 15. Radiology. 2005. PMID: 15601893
-
Dynamic chest radiography: flat-panel detector (FPD) based functional X-ray imaging.Radiol Phys Technol. 2016 Jul;9(2):139-53. doi: 10.1007/s12194-016-0361-6. Epub 2016 Jun 13. Radiol Phys Technol. 2016. PMID: 27294264 Review.
-
Dynamic Chest X-Ray Using a Flat-Panel Detector System: Technique and Applications.Korean J Radiol. 2021 Apr;22(4):634-651. doi: 10.3348/kjr.2020.1136. Epub 2020 Nov 30. Korean J Radiol. 2021. PMID: 33289365 Free PMC article. Review.
Cited by
-
Multi-reader multi-case studies using the area under the receiver operator characteristic curve as a measure of diagnostic accuracy: systematic review with a focus on quality of data reporting.PLoS One. 2014 Dec 26;9(12):e116018. doi: 10.1371/journal.pone.0116018. eCollection 2014. PLoS One. 2014. PMID: 25541977 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials