Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside study

Affiliations
Free article
Clinical Trial

Evaluating noninvasive ventilation using a monitoring system coupled to a ventilator: a bench-to-bedside study

C Rabec et al. Eur Respir J. 2009 Oct.
Free article

Abstract

Empirically determined noninvasive ventilation (NIV) settings may not achieve optimal ventilatory support. Some ventilators include monitoring modules to assess ventilatory quality. We conducted a bench-to-bedside study to assess the ventilatory quality of the VPAPIII-ResLink (ResMed, North Ryde, Australia). We tested the accuracy of minute ventilation (MV) and leak calculations given by VPAPIII-ResLink compared to those measured by a bench model at varied leak levels and ventilator settings. We systematically assessed NIV efficacy using this system from 2003 to 2006. Ventilation was considered inadequate if leak (>24 L x min(-1)), continuous desaturation (>30% of the trace) or desaturation dips (>3%) were present. On the bench test, both methods were highly correlated (r = 0.947, p>0.0001 and r = 0.959, p<0.0001 for leak and MV, respectively). We performed 222 assessments in 169 patients (aged 66.42+/-16 yrs, 100 males). Abnormalities were detected on 147 (66%) out of 222 occasions. Leak was the most common abnormality (34.2%) followed by desaturation dips (23.8%). The most effective therapeutic solutions were a chin strap if leak was detected (61.2%) and expiratory positive airway pressure increase for desaturation dips (59.5%). In 15.7% of cases, when abnormalities persisted, a polygraphy was performed. The systematic use of this device enables NIV to be optimised, limiting the indication of sleep studies to complex cases.