Toward submillimeter accuracy in the management of intrafraction motion: the integration of real-time internal position monitoring and multileaf collimator target tracking
- PMID: 19327907
- DOI: 10.1016/j.ijrobp.2008.12.057
Toward submillimeter accuracy in the management of intrafraction motion: the integration of real-time internal position monitoring and multileaf collimator target tracking
Abstract
Purpose: We report on an integrated system for real-time adaptive radiation delivery to moving tumors. The system combines two promising technologies-three-dimensional internal position monitoring using implanted electromagnetically excitable transponders and corresponding real-time beam adaptation using a dynamic multileaf collimator (DMLC).
Methods and materials: In a multi-institutional academic and industrial collaboration, a research version of the Calypso position monitoring system was integrated with a DMLC-based four-dimensional intensity-modulated radiotherapy delivery system using a Varian 120-leaf multileaf collimator (MLC). Two important determinants of system performance-latency (i.e., elapsed time between target motion and MLC response) and geometric accuracy-were investigated. Latency was quantified by acquiring continuous megavoltage X-ray images of a moving phantom (with embedded transponders) that was tracked in real time by a circular MLC field. The latency value was input into a motion prediction algorithm within the DMLC tracking system. Geometric accuracy was calculated as the root-mean-square positional error between the target and the centroid of the MLC aperture for patient-derived three-dimensional motion trajectories comprising two lung tumor traces and one prostate trace.
Results: System latency was determined to be approximately 220 milliseconds. Tracking accuracy was observed to be sub-2 mm for the respiratory motion traces and sub-1 mm for prostate motion.
Conclusion: We have developed and characterized a research version of a novel four-dimensional delivery system that integrates nonionizing radiation-based internal position monitoring and accurate real-time DMLC-based beam adaptation. This system represents a significant step toward achieving the eventual goal of geometrically ideal dose delivery to moving tumors.
Similar articles
-
Integration of real-time internal electromagnetic position monitoring coupled with dynamic multileaf collimator tracking: an intensity-modulated radiation therapy feasibility study.Int J Radiat Oncol Biol Phys. 2009 Jul 1;74(3):868-75. doi: 10.1016/j.ijrobp.2009.01.031. Epub 2009 Apr 23. Int J Radiat Oncol Biol Phys. 2009. PMID: 19394159
-
Implementation of a new method for dynamic multileaf collimator tracking of prostate motion in arc radiotherapy using a single kV imager.Int J Radiat Oncol Biol Phys. 2010 Mar 1;76(3):914-23. doi: 10.1016/j.ijrobp.2009.06.073. Epub 2009 Nov 10. Int J Radiat Oncol Biol Phys. 2010. PMID: 19910138
-
DMLC motion tracking of moving targets for intensity modulated arc therapy treatment: a feasibility study.Acta Oncol. 2009;48(2):245-50. doi: 10.1080/02841860802266722. Acta Oncol. 2009. PMID: 18720056
-
[Prostate localization systems for prostate radiotherapy].Cancer Radiother. 2006 Nov;10(6-7):394-401. doi: 10.1016/j.canrad.2006.09.002. Epub 2006 Oct 10. Cancer Radiother. 2006. PMID: 17035061 Review. French.
-
[A review of progress of real-time tumor tracking radiotherapy technology based on dynamic multi-leaf collimator].Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2017 Feb;34(1):145-9. doi: 10.7507/1001-5515.201607054. Sheng Wu Yi Xue Gong Cheng Xue Za Zhi. 2017. PMID: 29717603 Free PMC article. Review. Chinese.
Cited by
-
Detailed analysis of latencies in image-based dynamic MLC tracking.Med Phys. 2010 Sep;37(9):4998-5005. doi: 10.1118/1.3480504. Med Phys. 2010. PMID: 20964219 Free PMC article.
-
respiTrack: Patient-specific real-time respiratory tumor motion prediction using magnetic tracking.Int J Comput Assist Radiol Surg. 2020 Jun;15(6):953-962. doi: 10.1007/s11548-020-02174-3. Epub 2020 Apr 28. Int J Comput Assist Radiol Surg. 2020. PMID: 32347464 Free PMC article.
-
Monitoring tumor motion by real time 2D/3D registration during radiotherapy.Radiother Oncol. 2012 Feb;102(2):274-80. doi: 10.1016/j.radonc.2011.07.031. Epub 2011 Aug 30. Radiother Oncol. 2012. PMID: 21885144 Free PMC article.
-
Geometric characterization of an electromagnetic surface tracking system in a radiation therapy environment.J Appl Clin Med Phys. 2025 Jul;26(7):e70187. doi: 10.1002/acm2.70187. J Appl Clin Med Phys. 2025. PMID: 40660872 Free PMC article.
-
Accuracy assessment of wireless transponder tracking in the operating room environment.Int J Comput Assist Radiol Surg. 2018 Dec;13(12):1937-1948. doi: 10.1007/s11548-018-1838-z. Epub 2018 Aug 11. Int J Comput Assist Radiol Surg. 2018. PMID: 30099659
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources