Comparison of manual and automated measurements of the QT interval in healthy volunteers: an analysis of five thorough QT studies
- PMID: 19339965
- DOI: 10.1038/clpt.2009.34
Comparison of manual and automated measurements of the QT interval in healthy volunteers: an analysis of five thorough QT studies
Abstract
We analyzed five crossover, thorough QT (TQT) studies to compare automated, manual, and computer-assisted (CA) measurement methods. All the methods detected moxifloxacin-induced, baseline-adjusted, placebo-subtracted mean changes in Fridericia-corrected QT interval (QTcF), with peak effect ranging from 10 to 21 ms. The variability associated with manual and CA measurements was generally 5-28% greater than that associated with automated methods. The performances of automated, manual, and CA measurements were comparable for the purpose of demonstrating assay sensitivity in TQT studies with healthy volunteers.
Similar articles
-
Man versus machine: is there an optimal method for QT measurements in thorough QT studies?J Clin Pharmacol. 2006 Jun;46(6):598-612. doi: 10.1177/0091270006286900. J Clin Pharmacol. 2006. PMID: 16707406 Clinical Trial.
-
Improving the precision of QT measurements.Cardiol J. 2011;18(4):401-10. Cardiol J. 2011. PMID: 21769821 Clinical Trial.
-
A statistical assessment of QT data following placebo and moxifloxacin dosing in thorough QT studies.J Biopharm Stat. 2008;18(3):502-16. doi: 10.1080/10543400801995460. J Biopharm Stat. 2008. PMID: 18470759
-
Implications of methodological differences in digital electrocardiogram interval measurement.J Electrocardiol. 2006 Oct;39(4 Suppl):S152-6. doi: 10.1016/j.jelectrocard.2006.05.030. Epub 2006 Aug 21. J Electrocardiol. 2006. PMID: 16920141 Review.
-
Arrhythmogenic liability screening in cardiovascular safety pharmacology: commonality between non-clinical safety pharmacology and clinical thorough QT (TQT) studies.J Pharmacol Toxicol Methods. 2010 Sep-Oct;62(2):83-8. doi: 10.1016/j.vascn.2010.06.005. Epub 2010 Jun 19. J Pharmacol Toxicol Methods. 2010. PMID: 20601020 Review.
Cited by
-
The thorough QT/QTc study 4 years after the implementation of the ICH E14 guidance.Br J Pharmacol. 2010 Jan;159(1):49-57. doi: 10.1111/j.1476-5381.2009.00487.x. Epub 2009 Nov 18. Br J Pharmacol. 2010. PMID: 19922536 Free PMC article. Review.
-
Drug-induced QT interval prolongation: mechanisms and clinical management.Ther Adv Drug Saf. 2012 Oct;3(5):241-53. doi: 10.1177/2042098612454283. Ther Adv Drug Saf. 2012. PMID: 25083239 Free PMC article. Review.
-
Man versus Machine: Comparison of Automated and Manual Methodologies for Measuring the QTc Interval: A Prospective Study.Ann Noninvasive Electrocardiol. 2016 Jan;21(1):82-90. doi: 10.1111/anec.12277. Epub 2015 May 4. Ann Noninvasive Electrocardiol. 2016. PMID: 25944685 Free PMC article. Clinical Trial.
-
Effects of Dihydroartemisinin-Piperaquine Phosphate and Artemether-Lumefantrine on QTc Interval Prolongation.Sci Rep. 2019 Jan 28;9(1):777. doi: 10.1038/s41598-018-37112-6. Sci Rep. 2019. PMID: 30692558 Free PMC article.
-
Losmapimod concentration-QT relationship in healthy volunteers: meta-analysis of data from six clinical trials.Eur J Clin Pharmacol. 2013 Jun;69(6):1261-7. doi: 10.1007/s00228-012-1469-1. Epub 2013 Jan 17. Eur J Clin Pharmacol. 2013. PMID: 23325437
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources