Comparison of manual and automated measurements of the QT interval in healthy volunteers: an analysis of five thorough QT studies
- PMID: 19339965
- DOI: 10.1038/clpt.2009.34
Comparison of manual and automated measurements of the QT interval in healthy volunteers: an analysis of five thorough QT studies
Abstract
We analyzed five crossover, thorough QT (TQT) studies to compare automated, manual, and computer-assisted (CA) measurement methods. All the methods detected moxifloxacin-induced, baseline-adjusted, placebo-subtracted mean changes in Fridericia-corrected QT interval (QTcF), with peak effect ranging from 10 to 21 ms. The variability associated with manual and CA measurements was generally 5-28% greater than that associated with automated methods. The performances of automated, manual, and CA measurements were comparable for the purpose of demonstrating assay sensitivity in TQT studies with healthy volunteers.