Development of an automated blood sampling system for use in humans
- PMID: 19340690
- DOI: 10.1080/03091900802185970
Development of an automated blood sampling system for use in humans
Abstract
Many hormones are released in a pulsatile or burst-like pattern resulting in fluctuating blood levels that can undergo rapid modulation by physiological and pathological signals. To accurately measure these changes in hormone concentration requires frequent blood sampling, often over extended periods as the overall rhythmicity may vary over 24 hours. The aim of this study was to develop a computerized, automated blood sampling system which allows repeated stress-free blood sample collection from humans over an extended period under basal or test conditions. The system incorporates a peristaltic pump, fraction collector and standard infusion pump together with a custom built electronic control unit linked to a personal computer. Disposable tubing prevents cross-contamination between study participants. The computer programme is modifiable to adjust for the number of specimen tubes and volume of blood collected per sampling cycle. Patency of the collecting line is maintained with 0.9% saline, without the need for heparinization. To validate the system, 10-minute samples for cortisol were collected over 24 hours from five healthy volunteers, of whom two had additional concomitant ACTH sampling. Deconvolution analysis revealed an expected number of hormone secretory episodes and a non-pathological degree of orderliness within the data. There was high concordance between ACTH and cortisol secretory events. The ability of the system to allow multiple measurements and of the software program to link with other physiological monitoring equipment provides a powerful tool to study physiologic/pathophysiologic change in relation to blood hormone and other biomarker levels.
Similar articles
-
High-speed automated discrete blood sampling for positron emission tomography.J Nucl Med. 1993 Aug;34(8):1357-60. J Nucl Med. 1993. PMID: 8326399
-
Microfiltration sampling in rats and in cows: toward a portable device for continuous glucocorticoid hormone sampling.Analyst. 2010 Feb;135(2):390-6. doi: 10.1039/b921629d. Epub 2009 Nov 30. Analyst. 2010. PMID: 20098775
-
Rapid changes in chimpanzee (Pan troglodytes) urinary cortisol excretion.Horm Behav. 2004 Mar;45(3):209-13. doi: 10.1016/j.yhbeh.2003.09.015. Horm Behav. 2004. PMID: 15047016
-
Diffusion-sink sampling for minimally invasive measurement of tissue available hormones.Ann Biol Clin (Paris). 1993;51(10-11):911-5. Ann Biol Clin (Paris). 1993. PMID: 8210069 Review.
-
Cortisol assays and diagnostic laboratory procedures in human biological fluids.Clin Biochem. 2009 Aug;42(12):1205-17. doi: 10.1016/j.clinbiochem.2009.04.011. Epub 2009 May 4. Clin Biochem. 2009. PMID: 19414006 Review.
Cited by
-
Modeling pulsativity in the hypothalamic-pituitary-adrenal hormonal axis.Sci Rep. 2022 May 19;12(1):8480. doi: 10.1038/s41598-022-12513-w. Sci Rep. 2022. PMID: 35589935 Free PMC article.
-
Dynamics of ACTH-Mediated Regulation of Gene Transcription in ATC1 and ATC7 Adrenal Zona Fasciculata Cell Lines.Endocrinology. 2019 Mar 1;160(3):587-604. doi: 10.1210/en.2018-00840. Endocrinology. 2019. PMID: 30768667 Free PMC article.
-
Using Aptamers as a Novel Method for Determining GnRH/LH Pulsatility.Int J Mol Sci. 2020 Oct 7;21(19):7394. doi: 10.3390/ijms21197394. Int J Mol Sci. 2020. PMID: 33036411 Free PMC article. Review.
-
Circadian Clocks and the Interaction between Stress Axis and Adipose Function.Int J Endocrinol. 2015;2015:693204. doi: 10.1155/2015/693204. Epub 2015 Apr 27. Int J Endocrinol. 2015. PMID: 26000016 Free PMC article. Review.
-
Role of glucocorticoid negative feedback in the regulation of HPA axis pulsatility.Stress. 2018 Sep;21(5):403-416. doi: 10.1080/10253890.2018.1470238. Epub 2018 May 15. Stress. 2018. PMID: 29764284 Free PMC article. Review.
Publication types
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources