Detection of lipid core coronary plaques in autopsy specimens with a novel catheter-based near-infrared spectroscopy system
- PMID: 19356494
- DOI: 10.1016/j.jcmg.2008.06.001
Detection of lipid core coronary plaques in autopsy specimens with a novel catheter-based near-infrared spectroscopy system
Abstract
Objectives: This study sought to assess agreement between an intravascular near-infrared spectroscopy (NIRS) system and histology in coronary autopsy specimens.
Background: Lipid core plaques cannot be detected by conventional tests, yet are suspected to be the cause of most acute coronary syndromes. Near-infrared spectroscopy is widely used to determine the chemical content of substances. A NIRS system has been developed and used successfully in 99 patients.
Methods: Scanning NIRS was performed through blood in 212 coronary segments from 84 autopsy hearts. One histologic section was analyzed for every 2 mm of artery. Lipid core plaque of interest (LCP) was defined as a lipid core >60 degrees in circumferential extent, >200-microm thick, with a mean fibrous cap thickness <450 microm. The first 33 hearts were used to develop the algorithm; the subsequent 51 validation hearts were used in a prospective, double-blind manner to evaluate the accuracy of NIRS in detecting LCP. A NIRS-derived lipid core burden index for an entire artery was also validated by comparison to histologic findings.
Results: The LCPs were present in 115 of 2,649 (4.3%) sections from the 51 validation hearts. The algorithm prospectively identified LCP with a receiver-operator characteristic area of 0.80 (95% confidence interval [CI]: 0.76 to 0.85). The lipid core burden index detected the presence or absence of any fibroatheroma with an area under the curve of 0.86 (95% CI: 0.81 to 0.91). A retrospective analysis of lipid core burden index conducted in extreme artery segments with either no or extensive fibroatheroma yielded an area under the curve of 0.96 (95% CI: 0.92 to 1.00), confirming the accuracy of spectroscopy in identifying plaques with markedly different lipid content under ideal circumstances.
Conclusions: This novel catheter-based NIRS system accurately identified lipid core plaques through blood in a prospective study in coronary autopsy specimens. It is expected that this novel capability will be of assistance in the management of patients with coronary artery disease.
Comment in
-
Detection of vulnerable coronary artery plaques: lesion level risk assessment within the coronary arterial tree?JACC Cardiovasc Imaging. 2008 Sep;1(5):649-51. doi: 10.1016/j.jcmg.2008.07.003. JACC Cardiovasc Imaging. 2008. PMID: 19356495 No abstract available.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous