Comparison of QTinno, a fully automated electrocardiographic analysis program, to semiautomated electrocardiographic analysis methods in a drug safety study in healthy subjects
- PMID: 19362316
- DOI: 10.1016/j.jelectrocard.2009.02.001
Comparison of QTinno, a fully automated electrocardiographic analysis program, to semiautomated electrocardiographic analysis methods in a drug safety study in healthy subjects
Abstract
Background: Improved automated methods for electrocardiographic (ECG) analysis are needed, particularly for drug development purposes.
Objectives: This study compared a novel fully automated method for ECG analysis (QTinno; NewCardio, Santa Clara, CA) to 2 semiautomated digital methods: global measurement from the earliest QRS onset to the latest T-wave offset on representative superimposed beats (global) and tangent measurement on 3 consecutive beats in one lead (tangent).
Methods: All 3 methods were used to determine uncorrected and rate-corrected QT interval duration (QT and QTcF) and related metrics in 1422 digital 12-lead ECGs from a phase 1 drug study. Global and tangent annotations were manually adjusted by the same 3 cardiologists wherever necessary. No adjustments were made in QTinno determinations.
Results: QTinno returned QTcF change from time-matched baseline (DeltaQTcF) that differed minimally from both global and tangent methods (mean pairwise difference: 0.1 millisecond between QTinno and global, 1.1 milliseconds between QTinno and tangent). The average absolute QT and QTcF intervals by QTinno were approximately 5 milliseconds longer than global and 25 milliseconds longer than by tangent. QTinno had lower intrinsic variability for DeltaQTcF than either global or tangent (between-subject SD: QTinno 4.0 milliseconds, global 5.6 milliseconds, tangent 6.4 milliseconds; within-subject SD: QTinno 4.8 milliseconds, global 7.4 milliseconds, tangent 10.6 milliseconds). All methods were robust in detecting the largest placebo-adjusted mean time-matched DeltaQTcF (15-25 milliseconds) induced by study drug.
Conclusions: The methods show good agreement for drug-induced QTc prolongation. Lower intrinsic variability of DeltaQTcF by QTinno could facilitate smaller sample sizes or increase study power in thorough QTc studies.
Similar articles
-
Comparison of semiautomated and fully automated methods for QT measurement during a thorough QT/QTc study: variability and sample size considerations.J Clin Pharmacol. 2009 Aug;49(8):905-15. doi: 10.1177/0091270009337944. Epub 2009 Jun 19. J Clin Pharmacol. 2009. PMID: 19542314 Clinical Trial.
-
Semiautomated QT interval measurement in electrocardiograms from a thorough QT study: comparison of the grouped and ungrouped superimposed median beat methods.J Electrocardiol. 2012 May-Jun;45(3):225-30. doi: 10.1016/j.jelectrocard.2012.01.007. Epub 2012 Feb 22. J Electrocardiol. 2012. PMID: 22364647 Clinical Trial.
-
Intra- and interreader variability in QT interval measurement by tangent and threshold methods in a central electrocardiogram laboratory.J Electrocardiol. 2009 Jul-Aug;42(4):348-52. doi: 10.1016/j.jelectrocard.2009.01.003. Epub 2009 Mar 3. J Electrocardiol. 2009. PMID: 19261293
-
Implications of methodological differences in digital electrocardiogram interval measurement.J Electrocardiol. 2006 Oct;39(4 Suppl):S152-6. doi: 10.1016/j.jelectrocard.2006.05.030. Epub 2006 Aug 21. J Electrocardiol. 2006. PMID: 16920141 Review.
-
Drug-induced QRS morphology and duration changes.Cardiol J. 2008;15(6):505-9. Cardiol J. 2008. PMID: 19039753 Review.
Cited by
-
Drug-induced QT interval prolongation: mechanisms and clinical management.Ther Adv Drug Saf. 2012 Oct;3(5):241-53. doi: 10.1177/2042098612454283. Ther Adv Drug Saf. 2012. PMID: 25083239 Free PMC article. Review.
-
Detection of QT prolongation using a novel electrocardiographic analysis algorithm applying intelligent automation: prospective blinded evaluation using the Cardiac Safety Research Consortium electrocardiographic database.Am Heart J. 2012 Mar;163(3):365-71. doi: 10.1016/j.ahj.2011.11.009. Epub 2012 Feb 27. Am Heart J. 2012. PMID: 22424006 Free PMC article. Clinical Trial.
-
Man versus Machine: Comparison of Automated and Manual Methodologies for Measuring the QTc Interval: A Prospective Study.Ann Noninvasive Electrocardiol. 2016 Jan;21(1):82-90. doi: 10.1111/anec.12277. Epub 2015 May 4. Ann Noninvasive Electrocardiol. 2016. PMID: 25944685 Free PMC article. Clinical Trial.
-
The thorough QT/QTc study 4 years after the implementation of the ICH E14 guidance.Br J Pharmacol. 2010 Jan;159(1):49-57. doi: 10.1111/j.1476-5381.2009.00487.x. Epub 2009 Nov 18. Br J Pharmacol. 2010. PMID: 19922536 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources