Accuracy, precision and linearity of the portable flow-volume meter Microspiro HI-298
- PMID: 1936232
Accuracy, precision and linearity of the portable flow-volume meter Microspiro HI-298
Abstract
The accuracy, precision and linearity of a new portable flow-volume meter, the Microspiro HI-298 (Chest Corporation, Tokyo, Japan), was investigated using a Fleisch no. 4 pneumotachograph as a standard. After connection and calibration of the pneumotachograph and the Microspiro, a healthy subject performed 44 forced vital capacity (FVC) manoeuvres at different levels of lung inflation. The FVC of these expirations ranged from 2.5-5.1 l. Linear regression of Microspiro values (dependent variable) on Fleisch pneumotachograph values (independent variable) showed that a good linear relationship existed: Pearson correlation coefficients ranged from 0.938-0.985. Linearity of the Microspiro was good except for the peak expiratory flow rate (PEFR) and the maximal expiratory flow at 25% of the expired volume (MEF75). The random error (measure of precision) of all flow-volume (F-V) indices was lower than 5%. The systematic error (measure of accuracy) was low for the forced expiratory volume in one second (FEV1) and the FVC (1% and 4.6%, respectively) but much higher for the instantaneous expiratory flows (PEFR 11.0%; MEF75 7.0%; MEF50 8.5%; MEF25 11.4%). Only the total error in FEV1 complied with the tolerance of 4% of the European Community for Coal and Steel (ECCS). When the measured values were adjusted according to the regression equations of this study, all F-V indices were accurate and precise within 5%. It was concluded that the portable Microspiro HI-298 is a useful instrument for bedside, work-site spirometry and for use in general practice.(ABSTRACT TRUNCATED AT 250 WORDS)
Similar articles
-
[Evaluation of respiratory tract function in healthy women in the last month of uncomplicated pregnancy].Ann Acad Med Stetin. 2002;48:331-50. Ann Acad Med Stetin. 2002. PMID: 14601487 Clinical Trial. Polish.
-
Spirometric reference values from a Mediterranean population.Bull Eur Physiopathol Respir. 1986 May-Jun;22(3):217-24. Bull Eur Physiopathol Respir. 1986. PMID: 3730638
-
Regression equations for spirometry in children aged 6 to 17 years in Delhi region.Indian J Chest Dis Allied Sci. 2012 Jan-Mar;54(1):59-63. Indian J Chest Dis Allied Sci. 2012. PMID: 22779126
-
Ventilatory parameters in healthy nonsmoking adults of Adriatic islands (Yugoslavia).Eur Respir J. 1991 Sep;4(8):955-64. Eur Respir J. 1991. PMID: 1783087
-
The accuracy of a handheld portable spirometer.Chest. 1996 Jan;109(1):152-7. doi: 10.1378/chest.109.1.152. Chest. 1996. PMID: 8549178 Clinical Trial.
Cited by
-
Asthma and chronic bronchitis. Can family physicians predict rates of progression?Can Fam Physician. 1995 Nov;41:1868-76. Can Fam Physician. 1995. PMID: 8563504 Free PMC article. Clinical Trial.
-
Testing bronchial hyper-responsiveness: provocation or peak expiratory flow variability?Br J Gen Pract. 1997 Aug;47(421):487-92. Br J Gen Pract. 1997. PMID: 9302787 Free PMC article. Clinical Trial.
-
Diagnosis of asthma and chronic obstructive pulmonary disease in general practice.Br J Gen Pract. 1996 Mar;46(404):193-7. Br J Gen Pract. 1996. PMID: 8731629 Free PMC article.
-
Continuous and on demand use of bronchodilators in patients with non-steroid dependent asthma and chronic bronchitis: four-year follow-up randomized controlled study.Br J Gen Pract. 1995 May;45(394):239-44. Br J Gen Pract. 1995. PMID: 7619569 Free PMC article. Clinical Trial.
-
Evaluation of the suitability of weekly peak expiratory flow rate measurements in monitoring annual decline in lung function among patients with asthma and chronic bronchitis.Br J Gen Pract. 1996 Jan;46(402):15-8. Br J Gen Pract. 1996. PMID: 8745846 Free PMC article. Clinical Trial.