Diffusion-weighted imaging in patients with acute brain ischemia at 3 T: current possibilities and future perspectives comparing conventional echoplanar diffusion-weighted imaging and fast spin echo diffusion-weighted imaging sequences using BLADE (PROPELLER)
- PMID: 19363447
- DOI: 10.1097/RLI.0b013e3181a00d09
Diffusion-weighted imaging in patients with acute brain ischemia at 3 T: current possibilities and future perspectives comparing conventional echoplanar diffusion-weighted imaging and fast spin echo diffusion-weighted imaging sequences using BLADE (PROPELLER)
Abstract
Objectives: To compare diffusion-weighted imaging (DWI) based on a fast spin echo (FSE) sequence using BLADE (PROPELLER) with conventional DWI-echoplanar imaging (EPI) techniques at 3 T and to demonstrate the influence of hardware developments on signal-to-noise ratio (SNR) with these techniques using 12- and 32-channel head coils.
Materials and methods: Fourteen patients with brain ischemia were evaluated with DWI using EPI and FSE BLADE sequences, with a 12-channel head coil, in the axial plane and 1 additional plane (either sagittal or coronal). SNR and CNR were calculated from region-of-interest measurements. Scans were evaluated in a blinded fashion by 2 experienced neuroradiologists. SNR of both DWI techniques was evaluated in 12 healthy volunteers using different parallel imaging (PI) factors (for the EPI sequence) and both the 12- and 32-channel coils.
Results: DWI-BLADE sequences acquired with the 12-channel coil revealed a significant reduction in SNR (mean +/- SD) of ischemic lesions (SNR(lesion) [5.0 +/- 2.5]), normal brain (SNR(brain) [3.0 +/- 1.9]), and subsequently in CNR (3.0 +/- 1.8) as compared with the DWI-EPI sequence (SNR(lesion) [9.3 +/- 5.2], SNR(brain) [7.7 +/- 3.5], CNR [6.1 +/- 2.8], P < 0.001). Despite this reduction in SNR and CNR, the blinded read revealed a marked preference for the DWI-BLADE sequence, or equality between the sequences, in the majority of patients because lesion detection was degraded by susceptibility artifacts on axial DWI-EPI scans in 14% to 43% of cases (but in no instance with the DWI-BLADE sequence). In particular, preference for the DWI-BLADE sequence or equality between the 2 techniques for lesion detection in the brainstem and cerebellum was observed. On some DWI-BLADE scans, in the additional plane, radial-like artifacts degraded lesion detection.In volunteers, SNR was significantly improved using the 32-channel coil, irrespective of scan technique. Comparing DWI-EPI acquired with the 12-channel coil (iPAT = 2) to DWI-BLADE acquired with the 32-channel coil, comparable SNR values were obtained. The 32-channel coil also makes feasible, with DWI-EPI, an increase in the PI factor to 4, which allows for a further reduction of bulk susceptibility artifacts. However, still DWI-BLADE sequences performed better because of absence of bulk susceptibility artifacts at comparable SNR values.
Conclusion: Despite lower SNR at comparable PI factors, DWI-BLADE sequences acquired using the 12-channel coil are preferable in most instances, as compared with DWI-EPI sequences, because of the absence of susceptibility artifacts and subsequently improved depiction of ischemic lesions in the brainstem and cerebellum. With the 32-channel coil, recently FDA approved, DWI-BLADE acquired with an iPAT = 2 provides comparable SNR without bulk susceptibility artifacts as compared with the DWI-EPI sequences acquired for clinical routine to date and has the potential to replace the standard DWI technique for special indications like DWI of the cerebellum and the brainstem or in presence of metallic implants or hemorrhage.
Similar articles
-
Diffusion weighted imaging: a comprehensive evaluation of a fast spin echo DWI sequence with BLADE (PROPELLER) k-space sampling at 3 T, using a 32-channel head coil in acute brain ischemia.Invest Radiol. 2009 Oct;44(10):656-61. doi: 10.1097/RLI.0b013e3181af3f0e. Invest Radiol. 2009. PMID: 19724235
-
T1-weighted brain imaging with a 32-channel coil at 3T using TurboFLASH BLADE compared with standard cartesian k-space sampling.Invest Radiol. 2009 Mar;44(3):177-83. doi: 10.1097/RLI.0b013e318198a1fd. Invest Radiol. 2009. PMID: 19151605
-
Thin-section diffusion-weighted magnetic resonance imaging of the brain with parallel imaging.Acta Radiol. 2007 May;48(4):456-63. doi: 10.1080/02841850701297506. Acta Radiol. 2007. PMID: 17453529
-
[The clinical application of diffusion weighted magnetic resonance imaging to acute cerebrovascular disorders].No To Shinkei. 1998 Sep;50(9):787-95. No To Shinkei. 1998. PMID: 9789301 Review. Japanese.
-
Technical aspects of MR diffusion imaging of the body.Eur J Radiol. 2010 Dec;76(3):314-22. doi: 10.1016/j.ejrad.2010.02.018. Epub 2010 Mar 29. Eur J Radiol. 2010. PMID: 20299172 Review.
Cited by
-
Comparison of TGSE-BLADE DWI, RESOLVE DWI, and SS-EPI DWI in healthy volunteers and patients after cerebral aneurysm clipping.Sci Rep. 2022 Oct 21;12(1):17689. doi: 10.1038/s41598-022-22760-6. Sci Rep. 2022. PMID: 36271294 Free PMC article.
-
Hyperintense Vessels on T2-PROPELLER-FLAIR in Patients with Acute MCA Stroke: Prediction of Arterial Stenosis and Perfusion Abnormality.AJNR Am J Neuroradiol. 2015 Nov;36(11):2042-7. doi: 10.3174/ajnr.A4423. Epub 2015 Jul 30. AJNR Am J Neuroradiol. 2015. PMID: 26228881 Free PMC article.
-
Improved abdominal MRI in non-breath-holding children using a radial k-space sampling technique.Pediatr Radiol. 2015 Jun;45(6):840-6. doi: 10.1007/s00247-014-3244-1. Epub 2015 Jan 24. Pediatr Radiol. 2015. PMID: 25616364
-
EADC Values in Diagnosis of Renal Lesions by 3.0 T Diffusion-Weighted Magnetic Resonance Imaging: Compared with the ADC Values.Appl Magn Reson. 2013 Mar;44(3):349-363. doi: 10.1007/s00723-012-0376-z. Epub 2012 Jul 5. Appl Magn Reson. 2013. PMID: 23476099 Free PMC article.
-
Echo planar diffusion-weighted imaging: possibilities and considerations with 12- and 32-channel head coils.J Clin Imaging Sci. 2012;2:31. doi: 10.4103/2156-7514.96548. Epub 2012 May 23. J Clin Imaging Sci. 2012. PMID: 22754745 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous