Comparative study of four retentive anchor systems for implant supported overdentures--retention force changes
- PMID: 19371390
- DOI: 10.1111/j.1741-2358.2009.00286.x
Comparative study of four retentive anchor systems for implant supported overdentures--retention force changes
Abstract
Objectives: Wear of attachments leads to a loss of retention and potentially reduces the function of complete dentures. This study evaluated the retention force changes of different prefabricated attachment systems for implant-supported overdentures to estimate the wear constancy and applicability in clinical practice.
Methods: Four prefabricated attachment systems were tested [Group SG: retentive ball attachment (Straumann, Switzerland) with gold matrix, Group ST: retentive ball attachment (Straumann, Switzerland) with titanium spring matrix, Group IB: UNOR i-Ball with Ecco matrix (UNOR, Switzerland) and Group IMZ: IMZ-TwinPlus ball attachment with gold matrix (DENTSPLY Friadent, Germany)]. Ten samples of each system were subjected to 10,000 insertion-separation cycles.
Results: Results showed that all types of attachments showed wear, which led to a loss of retention force after an initial increase at the beginning of the wear simulation. Attachments with a plastic retention insert or gold matrices underwent the smallest changes in retention force. The titanium spring system showed the largest changes in retention force and a greater variation between the different cycles and specimen. This behaviour is probably caused by a large fitting tolerance of the titanium spring.
Conclusions: Attachment systems which possess a male and female component of different material composition are preferable. They show smaller changes in the retention force. For retention force increase and wear compensation, an attachment system should be adjustable.
Similar articles
-
Wear of ceramic and titanium ball attachments in subjects with an implant-retained overdenture: a controlled clinical trial.J Prosthet Dent. 2012 Feb;107(2):109-13. doi: 10.1016/S0022-3913(12)60035-3. J Prosthet Dent. 2012. PMID: 22304745 Clinical Trial.
-
Resistance to wear of four matrices with ball attachments for implant overdentures: a fatigue study.J Prosthodont. 2010 Dec;19(8):614-9. doi: 10.1111/j.1532-849X.2010.00613.x. J Prosthodont. 2010. PMID: 20546492
-
Effect of implant angulation upon retention of overdenture attachments.J Prosthodont. 2005 Mar;14(1):3-11. doi: 10.1111/j.1532-849X.2005.00005.x. J Prosthodont. 2005. PMID: 15733129
-
Attachment systems for mandibular two-implant overdentures: a review of in vitro investigations on retention and wear features.Int J Prosthodont. 2009 Sep-Oct;22(5):429-40. Int J Prosthodont. 2009. PMID: 20095190 Review.
-
Digital wear analysis and retention of poly-ether-ether-ketone retentive inserts versus conventional nylon inserts in locator retained mandibular overdentures: in-vitro study.Clin Oral Investig. 2024 Aug 6;28(9):468. doi: 10.1007/s00784-024-05831-y. Clin Oral Investig. 2024. PMID: 39105859 Free PMC article. Review.
Cited by
-
Wear of double crown systems: electroplated vs. casted female part.J Appl Oral Sci. 2012 May-Jun;20(3):384-91. doi: 10.1590/s1678-77572012000300015. J Appl Oral Sci. 2012. PMID: 22858708 Free PMC article.
-
Clinical retention force development of double crowns.Clin Oral Investig. 2012 Apr;16(2):407-11. doi: 10.1007/s00784-011-0530-2. Epub 2011 Mar 1. Clin Oral Investig. 2012. PMID: 21360104 Clinical Trial.
-
In vitro retention force measurement for three different attachment systems for implant-retained overdenture.J Indian Prosthodont Soc. 2016 Oct-Dec;16(4):380-385. doi: 10.4103/0972-4052.191284. J Indian Prosthodont Soc. 2016. PMID: 27746603 Free PMC article.
-
Effect of abutment angulation in the retention and durability of three overdenture attachment systems: An in vitro study.J Adv Prosthodont. 2016 Feb;8(1):21-9. doi: 10.4047/jap.2016.8.1.21. Epub 2016 Feb 23. J Adv Prosthodont. 2016. PMID: 26949484 Free PMC article.
-
Retention forces in mini-dental-implant retained mandibular overdentures: 10-year outcomes of a non-comparative longitudinal observational study.Int J Implant Dent. 2025 Apr 17;11(1):32. doi: 10.1186/s40729-025-00620-y. Int J Implant Dent. 2025. PMID: 40244532 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources