Comparative study of four retentive anchor systems for implant supported overdentures--retention force changes

Affiliations
Comparative Study

Comparative study of four retentive anchor systems for implant supported overdentures--retention force changes

Stefan Bayer et al. Gerodontology. 2009 Dec.

Abstract

Objectives: Wear of attachments leads to a loss of retention and potentially reduces the function of complete dentures. This study evaluated the retention force changes of different prefabricated attachment systems for implant-supported overdentures to estimate the wear constancy and applicability in clinical practice.

Methods: Four prefabricated attachment systems were tested [Group SG: retentive ball attachment (Straumann, Switzerland) with gold matrix, Group ST: retentive ball attachment (Straumann, Switzerland) with titanium spring matrix, Group IB: UNOR i-Ball with Ecco matrix (UNOR, Switzerland) and Group IMZ: IMZ-TwinPlus ball attachment with gold matrix (DENTSPLY Friadent, Germany)]. Ten samples of each system were subjected to 10,000 insertion-separation cycles.

Results: Results showed that all types of attachments showed wear, which led to a loss of retention force after an initial increase at the beginning of the wear simulation. Attachments with a plastic retention insert or gold matrices underwent the smallest changes in retention force. The titanium spring system showed the largest changes in retention force and a greater variation between the different cycles and specimen. This behaviour is probably caused by a large fitting tolerance of the titanium spring.

Conclusions: Attachment systems which possess a male and female component of different material composition are preferable. They show smaller changes in the retention force. For retention force increase and wear compensation, an attachment system should be adjustable.