. 2009 Jun;55(6):1092-9.
doi: 10.1373/clinchem.2008.120717. Epub 2009 Apr 16.

Measurement and quality control issues in multiplex protein assays: a case study

Affiliations

Measurement and quality control issues in multiplex protein assays: a case study

Allison A Ellington et al. Clin Chem. 2009 Jun.

Abstract

Background: Multiplex arrays are increasingly used for measuring protein biomarkers. Advantages of this approach include specimen conservation, limited sample handling, and decreased time and cost, but the challenges of optimizing assay format for each protein, selecting common dilution factors, and establishing robust quality control algorithms are substantial. Here, we use measurements of 15 protein biomarkers from a large study to illustrate processing, analytic, and quality control issues with multiplexed immunoassays.

Methods: We contracted with ThermoScientific for duplicate measurements of 15 proteins in 2322 participants from a community-based cohort, a plasma control, and recombinant protein controls using 2 custom planar microarrays with 6 (panel A) or 9 (panel B) capture antibodies printed in each well. We selected constituent analytes in each panel based on endogenous concentrations and assay availability. Protocols were standardized for sample processing, storage, and freeze-thaw exposures. We analyzed data for effects of deviations from processing protocols, precision, and bias.

Results: Measurements were within reportable ranges for each of the assays; however, concentrations for 7 of the 15 proteins were not centered on the dose-response curves. An additional freeze-thaw cycle and erroneous sample dilution for a subset of samples produced significantly different results. Measurements with large differences between duplicates were seen to cluster by analyte, plate, and participant. Conventional univariate quality control algorithms rejected many plates. Plate-specific medians of cohort and plasma control data significantly covaried, an observation important for development of alternative quality control algorithms.

Conclusions: Multiplex measurements present difficult challenges that require further analytical and statistical developments.

PubMed Disclaimer

Figures

Fig. 1
Fig. 1. Representative dose–response curves
Vertical dotted lines demark the 5th and 95th percentiles of cohort data. Arrows, squares, triangles, and diamonds depict the median values of the plasma control, recombinant protein quality control 1 (QC1), recombinant protein QC2, and recombinant protein QC3, respectively.
Fig. 2
Fig. 2. (A), (B), and (C) illustrate the covariance of plate-specific population-normalized cohort means and plasma control means across 66 microplates for 3 representative proteins

Comment in

  • Quality control and protein microarrays.
    Kricka LJ, Master SR. Kricka LJ, et al. Clin Chem. 2009 Jun;55(6):1053-5. doi: 10.1373/clinchem.2009.126557. Epub 2009 Apr 24. Clin Chem. 2009. PMID: 19395435 No abstract available.

References

    1. MAQC Consortium The MicroArray Quality Control (MAQC) project shows interand intraplatform reproducibility of gene expression measurements. Nat Biotechnol. 2006;24:1151–61. - PMC - PubMed
    1. van’t Veer LJDH, van de Vijver MJ, He YD, Hart AA, Mao M, Peterse HL, et al. Gene expression profiling predicts clinical outcome of breast cancer. Nature (Lond) 2002;415:530–6. - PubMed
    1. Rifai N, Gillette MA, Carr SA. Protein biomarker discovery and validation: the long and uncertain path to clinical utility. Nat Biotechnol. 2006;24:971–83. - PubMed
    1. Schweitzer B, Kingsmore SF. Measuring proteins on microarrays. Curr Opin Biotechnol. 2002;13:14–9. - PubMed
    1. Endogen SearchLight. [Accessed November 2008]. http://www.endogen.com/services/

Publication types