List mode-driven cardiac and respiratory gating in PET
- PMID: 19372491
- DOI: 10.2967/jnumed.108.059204
List mode-driven cardiac and respiratory gating in PET
Abstract
Gating methods acquiring biosignals (such as electrocardiography [ECG] and respiration) during PET enable one to reduce motion effects that potentially lead to image blurring and artifacts. This study evaluated different cardiac and respiratory gating methods: one based on ECG signals for cardiac gating and video signals for respiratory gating; 2 others based on measured inherent list mode events.
Methods: Twenty-nine patients with coronary artery disease underwent a 20-min ECG-gated single-bed list mode PET scan of the heart. Of these, 17 were monitored by a video camera registering a marker on the patient's abdomen, thus capturing the respiratory motion for PET gating (video method). Additionally, respiratory and cardiac gating information was deduced without auxiliary measurements by dividing the list mode stream in 50-ms frames and then either determining the number of coincidences (sensitivity method) or computing the axial center of mass and SD of the measured counting rates in the same frames (center-of-mass method). The gated datasets (respiratory and cardiac gating) were reconstructed without attenuation correction. Measured wall thicknesses, maximum displacement of the left ventricular wall, and ejection fraction served as measures of the exactness of gating.
Results: All methods successfully captured respiratory motion and significantly decreased motion-induced blurring in the gated images. The center-of-mass method resulted in significantly larger left ventricular wall displacements than did the sensitivity method (P < 0.02); other differences were nonsignificant. List mode-based cardiac gating was found to work well for patients with high (18)F-FDG uptake when the center-of-mass method was used, leading to an ejection fraction correlation coefficient of r = 0.95 as compared with ECG-based gating. However, the sensitivity method did not always result in valid cardiac gating information, even in patients with high (18)F-FDG uptake.
Conclusion: Our study demonstrated that valid gating signals during PET scans cannot be obtained only by tracking the external motion or applying an ECG but also by simply analyzing the PET list mode stream on a frame-by-frame basis.
Similar articles
-
Detection of respiratory tumour motion using intrinsic list mode-driven gating in positron emission tomography.Eur J Nucl Med Mol Imaging. 2010 Dec;37(12):2315-27. doi: 10.1007/s00259-010-1533-y. Epub 2010 Jul 6. Eur J Nucl Med Mol Imaging. 2010. PMID: 20607534
-
Erroneous cardiac ECG-gated PET list-mode trigger events can be retrospectively identified and replaced by an offline reprocessing approach: first results in rodents.Phys Med Biol. 2013 Nov 21;58(22):7937-59. doi: 10.1088/0031-9155/58/22/7937. Epub 2013 Oct 29. Phys Med Biol. 2013. PMID: 24165267
-
Data-driven gating in PET: Influence of respiratory signal noise on motion resolution.Med Phys. 2018 Jul;45(7):3205-3213. doi: 10.1002/mp.12987. Epub 2018 Jun 8. Med Phys. 2018. PMID: 29782653
-
Added Value of Respiratory Gating in Positron Emission Tomography for the Clinical Management of Lung Cancer Patients.Semin Nucl Med. 2022 Nov;52(6):745-758. doi: 10.1053/j.semnuclmed.2022.04.006. Epub 2022 May 25. Semin Nucl Med. 2022. PMID: 35643531 Review.
-
Gating Approaches in Cardiac PET Imaging.PET Clin. 2019 Apr;14(2):271-279. doi: 10.1016/j.cpet.2018.12.008. Epub 2019 Feb 1. PET Clin. 2019. PMID: 30826024 Free PMC article. Review.
Cited by
-
Imaging moving heart structures with PET.J Nucl Cardiol. 2016 Jun;23(3):486-90. doi: 10.1007/s12350-015-0094-7. Epub 2015 Mar 26. J Nucl Cardiol. 2016. PMID: 25809083 No abstract available.
-
Evaluation of the endothelial function in hypertensive patients with 13N-ammonia PET.J Nucl Cardiol. 2012 Oct;19(5):979-86. doi: 10.1007/s12350-012-9584-z. Epub 2012 Jun 12. J Nucl Cardiol. 2012. PMID: 22689073
-
Motion detection and amelioration in a dedicated cardiac solid-state CZT SPECT device.Med Biol Eng Comput. 2017 Apr;55(4):663-671. doi: 10.1007/s11517-016-1548-z. Epub 2016 Jul 14. Med Biol Eng Comput. 2017. PMID: 27417733
-
(18)F-fluorodeoxyglucose PET imaging of coronary atherosclerosis and plaque inflammation.Curr Cardiol Rep. 2010 Mar;12(2):179-84. doi: 10.1007/s11886-010-0095-8. Curr Cardiol Rep. 2010. PMID: 20425174
-
Optimizing bioimpedance measurement configuration for dual-gated nuclear medicine imaging: a sensitivity study.Med Biol Eng Comput. 2011 Jul;49(7):783-91. doi: 10.1007/s11517-011-0787-2. Epub 2011 May 27. Med Biol Eng Comput. 2011. PMID: 21618055
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical