In vitro comparison of two generations of Licox and Neurotrend catheters

Affiliations
Comparative Study

In vitro comparison of two generations of Licox and Neurotrend catheters

Iain Haitsma et al. Acta Neurochir Suppl. 2008.

Abstract

Background: Clinical reports on brain tissue oxygen tension differ in suggested threshold values for defining cerebral ischemia using the Licox and Neurotrend/Paratrend system. We evaluated in vitro performance of both first and second generation devices.

Materials and methods: Response rate and accuracy in solutions with oxygen tensions from 0 to 150 mm Hg were measured.

Findings: Ninety-five percent Response times were 102 +/- 13 seconds for first generation Licox probes and 135 +/- 24 s for Paratrend (n = 6, each probe), with second generation probes at 134 +/- 4 and 116 +/- 16 s respectively. At pO2 150 mmHg Licox and Paratrend probes were accurate with 2.2% and 2.1% error, respectively and 2.6% and 4.1% for later generation. At pO2 18 mmHg, Paratrend overestimated by 16.5% (absolute error range 2.18 to 4.18 mmHg), 7.4% for Neurotrend, Licox underestimated by 1.8% (absolute error range 0.08 to 0.52 mmHg) with 3.6% for the second generation probe.

Conclusions: Differences between the first generation probe types, while statistically significant (p < 0.001), may not be clinically relevant. Overestimation of pO2 by Neurotrend and small underestimation by Licox partially explain differences in published thresholds for cerebral ischemia. The Neurotrend was slightly more accurate and faster than the Paratrend system.