. 2009 Apr 25;373(9673):1440-6.
doi: 10.1016/S0140-6736(09)60248-8.

Effectiveness of haemodialysis access with an autologous tissue-engineered vascular graft: a multicentre cohort study

Affiliations
Clinical Trial

Effectiveness of haemodialysis access with an autologous tissue-engineered vascular graft: a multicentre cohort study

Todd N McAllister et al. Lancet. .

Abstract

Background: Application of a tissue-engineered vascular graft for small-diameter vascular reconstruction has been a long awaited and much anticipated advance for vascular surgery. We report results after a minimum of 6 months of follow-up for the first ten patients implanted with a completely biological and autologous tissue-engineered vascular graft.

Methods: Ten patients with end-stage renal disease who had been receiving haemodialysis through an access graft that had a high probability of failure, and had had at least one previous access failure, were enrolled from centres in Argentina and Poland between September, 2004, and April, 2007. Completely autologous tissue-engineered vascular grafts were grown in culture supplemented with bovine serum, implanted as arteriovenous shunts, and assessed for both mechanical stability during the safety phase (0-3 months) and effectiveness after haemodialysis was started.

Findings: Three grafts failed within the safety phase, which is consistent with failure rates expected for this high-risk patient population. One patient was withdrawn from the study because of severe gastrointestinal bleeding shortly before implantation, and another died of unrelated causes during the safety period with a patent graft. The remaining five patients had grafts functioning for haemodialysis 6-20 months after implantation, and a total of 68 patient-months of patency. In these five patients, only one intervention (surgical correction) was needed to maintain secondary patency. Overall, primary patency was maintained in seven (78%) of the remaining nine patients 1 month after implantation and five (60%) of the remaining eight patients 6 months after implantation.

Interpretation: Our proportion of primary patency in this high-risk cohort approaches Dialysis Outcomes Quality Initiative objectives (76% of patients 3 months after implantation) for arteriovenous fistulas, averaged across all patient populations.

PubMed Disclaimer

Comment in

Publication types