Intraoperative computed tomography with integrated navigation system in a multidisciplinary operating suite
- PMID: 19404103
- DOI: 10.1227/01.NEU.0000340785.51492.B5
Intraoperative computed tomography with integrated navigation system in a multidisciplinary operating suite
Abstract
Objective: We report our preliminary experience in a prospective series of patients with regard to feasibility, work flow, and image quality using a multislice computed tomographic (CT) scanner combined with a frameless neuronavigation system (NNS).
Methods: A sliding gantry 40-slice CT scanner was installed in a preexisting operating room. The scanner was connected to a frameless infrared-based NNS. Image data was transferred directly from the scanner into the navigation system. This allowed updating of the NNS during surgery by automated image registration based on the position of the gantry. Intraoperative CT angiography was possible. The patient was positioned on a radiolucent operating table that fits within the bore of the gantry. During image acquisition, the gantry moved over the patient. This table allowed all positions and movements like any normal operating table without compromising the positioning of the patient. For cranial surgery, a carbon-made radiolucent head clamp was fixed to the table.
Results: Experience with the first 230 patients confirms the feasibility of intraoperative CT scanning (136 patients with intracranial pathology, 94 patients with spinal lesions). After a specific work flow, interruption of surgery for intraoperative scanning can be limited to 10 to 15 minutes in cranial surgery and to 9 minutes in spinal surgery. Intraoperative imaging changed the course of surgery in 16 of the 230 cases either because control CT scans showed suboptimal screw position (17 of 307 screws, with 9 in 7 patients requiring correction) or that tumor resection was insufficient (9 cases). Intraoperative CT angiography has been performed in 7 cases so far with good image quality to determine residual flow in an aneurysm. Image quality was excellent in spinal and cranial base surgery.
Conclusion: The system can be installed in a preexisting operating environment without the need for special surgical instruments. It increases the safety of the patient and the surgeon without necessitating a change in the existing surgical protocol and work flow. Imaging and updating of the NNS can be performed at any time during surgery with very limited time and modification of the surgical setup. Multidisciplinary use increases utilization of the system and thus improves the cost-efficiency relationship.
Similar articles
-
Intraoperative computed tomography with integrated navigation system in spinal stabilizations.Spine (Phila Pa 1976). 2009 Dec 15;34(26):2919-26. doi: 10.1097/BRS.0b013e3181b77b19. Spine (Phila Pa 1976). 2009. PMID: 20010400
-
Intraoperative computed tomography with an integrated navigation system in stabilization surgery for complex craniovertebral junction malformation.J Spinal Disord Tech. 2014 Jul;27(5):245-52. doi: 10.1097/BSD.0b013e3182621900. J Spinal Disord Tech. 2014. PMID: 22688615 Clinical Trial.
-
Intraoperative computed tomography guided neuronavigation: concepts, efficiency, and work flow.Comput Aided Surg. 1998;3(4):174-82. doi: 10.1002/(SICI)1097-0150(1998)3:4<174::AID-IGS7>3.0.CO;2-G. Comput Aided Surg. 1998. PMID: 10027491
-
Updating of neuronavigation based on images intraoperatively acquired with a mobile computerized tomographic scanner: technical note.Minim Invasive Neurosurg. 2003 Apr;46(2):117-20. doi: 10.1055/s-2003-39344. Minim Invasive Neurosurg. 2003. PMID: 12761685 Review.
-
Intraoperative spinal navigation.Spine (Phila Pa 1976). 2003 Aug 1;28(15 Suppl):S54-61. doi: 10.1097/01.BRS.0000076899.78522.D9. Spine (Phila Pa 1976). 2003. PMID: 12897475 Review.
Cited by
-
Mobile computed tomography : three year clinical experience in Korea.J Korean Neurosurg Soc. 2013 Jan;53(1):39-42. doi: 10.3340/jkns.2013.53.1.39. Epub 2013 Jan 31. J Korean Neurosurg Soc. 2013. PMID: 23440939 Free PMC article.
-
Minimizing OCT quantification error via a surface-tracking imaging probe.Biomed Opt Express. 2021 Jun 10;12(7):3992-4002. doi: 10.1364/BOE.423233. eCollection 2021 Jul 1. Biomed Opt Express. 2021. PMID: 34457394 Free PMC article.
-
Top 100 Most Cited Articles on Intraoperative Image-Guided Navigation in Spine Surgery.Cureus. 2024 Aug 27;16(8):e67950. doi: 10.7759/cureus.67950. eCollection 2024 Aug. Cureus. 2024. PMID: 39328685 Free PMC article. Review.
-
Estimation of brain deformation for volumetric image updating in protoporphyrin IX fluorescence-guided resection.Stereotact Funct Neurosurg. 2010;88(1):1-10. doi: 10.1159/000258143. Epub 2009 Nov 12. Stereotact Funct Neurosurg. 2010. PMID: 19907205 Free PMC article.
-
Detection of human brain cancer infiltration ex vivo and in vivo using quantitative optical coherence tomography.Sci Transl Med. 2015 Jun 17;7(292):292ra100. doi: 10.1126/scitranslmed.3010611. Sci Transl Med. 2015. PMID: 26084803 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical