Effect of MLC leaf width and PTV margin on the treatment planning of intensity-modulated stereotactic radiosurgery (IMSRS) or radiotherapy (IMSRT)
- PMID: 19410139
- DOI: 10.1016/j.meddos.2008.06.002
Effect of MLC leaf width and PTV margin on the treatment planning of intensity-modulated stereotactic radiosurgery (IMSRS) or radiotherapy (IMSRT)
Abstract
We studied the effect of MLC (multileaf collimator) leaf width and PTV (planning target volume) margin on treatment planning of intensity modulated stereotactic radiosurgery (IMSRS) or radiotherapy (IMSRT). Twelve patients previously treated with IMSRS/IMSRT were retrospectively planned with 5- and 3-mm MLC leaf widths and 3- and 2-mm PTV margins using the already contoured clinical target volume and critical structures. The same beam arrangement, planning parameters, and optimization method were used in each of the 4 plans for a given patient. Each plan was normalized so that the prescription dose covered at least 99% of the PTV. Plan indices--D(mean) (mean dose), conformity index (CI), V(70) (volume receiving >or= 70% of the prescription dose), and V(50) (volume receiving >or= 50% of the prescription dose)--were calculated from the dose-volume histograms (DVHs) of the PTV, normal tissue, and organs at risk (OARs). Hypothesis testing was performed on the mean ratios of plan indices to determine the statistical significance of the relative differences. The PTV was well covered for all plans, as no significant differences were observed for D(95), V(95), D(max), D(min), and D(mean) of the PTV. The irradiated volume was approximately 23% smaller when 2-mm instead of 3-mm PTV margin was used, but it was only reduced by approximately 6% when the MLC leaf width was reduced from 5 mm to 3 mm. For normal tissue and brainstem, V(70), V(50), and D(mean) were reduced more effectively by a decrease in MLC width, while D(mean) of optic nerve and chiasm were more sensitive to a change in PTV margin. The DVH statistics for the PTV and normal structures from the treatment plan with 5-mm MLC and 2-mm PTV margin were equal to those with 3-mm MLC and 3-mm PTV margin. PTV margin reduction is more effective in sparing the normal tissue and OARs than a reduction in MLC leaf width. For IMSRS, where highly accurate setup and small PTV margins are routinely employed, the use of 5-mm MLC is therefore less desirable.
Similar articles
-
Impact of IMRT and leaf width on stereotactic body radiotherapy of liver and lung lesions.Int J Radiat Oncol Biol Phys. 2005 Apr 1;61(5):1572-81. doi: 10.1016/j.ijrobp.2004.12.075. Int J Radiat Oncol Biol Phys. 2005. PMID: 15817364
-
Intensity modulated radiation therapy with multileaf collimators of different leaf widths: a comparison of achievable dose distributions.Radiother Oncol. 2005 Apr;75(1):106-11. doi: 10.1016/j.radonc.2005.02.007. Epub 2005 Apr 18. Radiother Oncol. 2005. PMID: 15878108 Clinical Trial.
-
Does intensity-modulated stereotactic radiotherapy achieve superior target conformity than conventional stereotactic radiotherapy in different intracranial tumours?Clin Oncol (R Coll Radiol). 2009 Jun;21(5):408-16. doi: 10.1016/j.clon.2009.02.002. Epub 2009 Mar 5. Clin Oncol (R Coll Radiol). 2009. PMID: 19268555
-
Optical tracking technology in stereotactic radiation therapy.Med Dosim. 2007 Summer;32(2):111-20. doi: 10.1016/j.meddos.2007.01.008. Med Dosim. 2007. PMID: 17472890 Review.
-
Dose selection in stereotactic radiosurgery.Neurosurg Clin N Am. 1999 Apr;10(2):271-80. Neurosurg Clin N Am. 1999. PMID: 10099092 Review.
Cited by
-
Restricted single isocenter for multiple targets dynamic conformal arc (RSIMT DCA) technique for brain stereotactic radiosurgery (SRS) planning.J Radiosurg SBRT. 2018;5(2):145-156. J Radiosurg SBRT. 2018. PMID: 29657895 Free PMC article.
-
Optimization of Minimum Segment Width Parameter in the Intensity-Modulated Radiotherapy Plan for Esophageal Cancer.Int J Gen Med. 2021 Dec 16;14:9913-9921. doi: 10.2147/IJGM.S336269. eCollection 2021. Int J Gen Med. 2021. PMID: 34938110 Free PMC article.
-
Spine SBRT With Halcyon™: Plan Quality, Modulation Complexity, Delivery Accuracy, and Speed.Front Oncol. 2019 Apr 26;9:319. doi: 10.3389/fonc.2019.00319. eCollection 2019. Front Oncol. 2019. PMID: 31106151 Free PMC article.
-
Dosimetric comparison of 2.5 mm vs. 3.0 mm leaf width micro-multileaf collimator-based treatment systems for intracranial stereotactic radiosurgery using dynamic conformal arcs: implications for treatment planning.Jpn J Radiol. 2011 Nov;29(9):630-8. doi: 10.1007/s11604-011-0606-6. Epub 2011 Sep 29. Jpn J Radiol. 2011. PMID: 21956368
-
Dosimetric sensitivity of leaf width on volumetric modulated arc therapy plan quality: an objective approach.Rep Pract Oncol Radiother. 2022 Mar 22;27(1):76-85. doi: 10.5603/RPOR.a2022.0001. eCollection 2022. Rep Pract Oncol Radiother. 2022. PMID: 35402026 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical