Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays
- PMID: 19420172
- PMCID: PMC2708481
- DOI: 10.1128/JCM.01761-08
Evaluation of the Abbott m2000 RealTime human immunodeficiency virus type 1 (HIV-1) assay for HIV load monitoring in South Africa compared to the Roche Cobas AmpliPrep-Cobas Amplicor, Roche Cobas AmpliPrep-Cobas TaqMan HIV-1, and BioMerieux NucliSENS EasyQ HIV-1 assays
Abstract
The implementation of antiretroviral therapy demands the need for increased access to viral load (VL) monitoring. Newer real-time VL testing technologies are faster and have larger dynamic ranges and fully automated extraction to benefit higher throughputs in resource-poor environments. The Abbott RealTime human immunodeficiency virus type 1 (HIV-1) assay was evaluated as a new option for testing for HIV-1 subtype C in South Africa, and its performance was compared to the performance of existing assays (the Cobas AmpliPrep-Cobas TaqMan HIV-1, version 1, assay; the AmpliPrep-Cobas Monitor standard HIV-1 assay; and the NucliSENS EasyQ-EasyMag HIV-1 assay) in a high-throughput laboratory. The total precision of the RealTime HIV-1 assay was acceptable over all viral load ranges. This assay compared most favorably with the Cobas AmpliPrep-Cobas TaqMan HIV-1 assay (R(2) = 0.904), with a low standard deviation of difference being detected (0.323 copies/ml). The bias against comparator assays ranged from -0.001 copies/ml to -0.228 copies/ml. Variability in the reporting of VLs for a 20-member subtype panel compared to the variability of other assays was noted with subtypes G and CRF02-AG. The RealTime HIV-1 assay can test 93 samples per day with minimal manual preparation, less staff, and the minimization of contamination through automation. This assay is suitable for HIV-1 subtype C VL quantification in South Africa.
Figures
References
-
- Alaeus, A., E. Lilja, S. Herman, J. Spadoro, J. Wang, and J. Albert. 1999. Assay of plasma samples representing different HIV-1 genetic subtypes: an evaluation of new versions of the Amplicor HIV-1 monitor assay. AIDS Res. Hum Retrovir. 15889-894. - PubMed
-
- Bekker, L. G., L. Myer, C. Orrell, S. Lawn, and R. Wood. 2006. Rapid scale-up of a community-based HIV treatment service: programme performance over 3 consecutive years in Guguletu, South Africa. S. Afr. Med. J. 96315-320. - PubMed
-
- Bland, J. M., and D. G. Altman. 1999. Measuring agreement in method comparison studies. Stat. Methods Med. Res. 8135-160. - PubMed
-
- Brambilla, D., S. Granger, and J. Bremer. 2000. Abstr. 7th Conf. Retrovir. Opportunist. Infect., abstr. 774.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical