A novel method for contrast-to-noise ratio (CNR) evaluation of digital mammography detectors

Affiliations

A novel method for contrast-to-noise ratio (CNR) evaluation of digital mammography detectors

P Baldelli et al. Eur Radiol. 2009 Sep.

Abstract

The purpose of this study was to test a new, simple method of evaluating the contrast-to-noise ratio (CNR) over the entire image field of a digital detector and to compare different mammography systems. Images were taken under clinical exposure conditions for a range of simulated breast thicknesses using poly(methyl methacrylate) (PMMA). At each PMMA thickness, a second image which included an additional 0.2-mm Al sheet was also acquired. Image processing software was used to calculate the CNR in multiple regions of interest (ROI) covering the entire area of the detector in order to obtain a 'CNR image'. Five detector types were evaluated, two CsI-alphaSi (GE Healthcare) flat panel systems, one alphaSe (Hologic) flat panel system and a two generations of scanning photon counting digital detectors (Sectra). Flat panel detectors exhibit better CNR uniformity compared with the first-generation scanning photon counting detector in terms of mean pixel value variation. However, significant improvement in CNR uniformity was observed for the next-generation scanning detector. The method proposed produces a map of the CNR and a measurement of uniformity throughout the entire image field of the detector. The application of this method enables quality control measurement of individual detectors and a comparison of detectors using different technologies.

PubMed Disclaimer

References

    1. Med Phys. 2000 Oct;27(10):2408-16 - PubMed
    1. Int J Gynaecol Obstet. 2003 Sep;82(3):319-26 - PubMed
    1. Cancer. 1994 Jul 1;74(1 Suppl):311-22 - PubMed
    1. Med Phys. 2006 Nov;33(11):4221-35 - PubMed
    1. Br J Radiol. 2000 Oct;73(874):1056-67 - PubMed

Publication types

LinkOut - more resources