Bilevel vs ICU ventilators providing noninvasive ventilation: effect of system leaks: a COPD lung model comparison
- PMID: 19429723
- DOI: 10.1378/chest.08-3018
Bilevel vs ICU ventilators providing noninvasive ventilation: effect of system leaks: a COPD lung model comparison
Abstract
Background: Noninvasive positive-pressure ventilation (NPPV) modes are currently available on bilevel and ICU ventilators. However, little data comparing the performance of the NPPV modes on these ventilators are available.
Methods: In an experimental bench study, the ability of nine ICU ventilators to function in the presence of leaks was compared with a bilevel ventilator using the IngMar ASL5000 lung simulator (IngMar Medical; Pittsburgh, PA) set at a compliance of 60 mL/cm H(2)O, an inspiratory resistance of 10 cm H(2)O/L/s, an expiratory resistance of 20 cm H(2)O/ L/s, and a respiratory rate of 15 breaths/min. All of the ventilators were set at 12 cm H(2)O pressure support and 5 cm H(2)O positive end-expiratory pressure. The data were collected at baseline and at three customized leaks.
Main results: At baseline, all of the ventilators were able to deliver adequate tidal volumes, to maintain airway pressure, and to synchronize with the simulator, without missed efforts or auto-triggering. As the leak was increased, all of the ventilators (except the Vision [Respironics; Murrysville, PA] and Servo I [Maquet; Solna, Sweden]) needed adjustment of sensitivity or cycling criteria to maintain adequate ventilation, and some transitioned to backup ventilation. Significant differences in triggering and cycling were observed between the Servo I and the Vision ventilators.
Conclusions: The Vision and Servo I were the only ventilators that required no adjustments as they adapted to increasing leaks. There were differences in performance between these two ventilators, although the clinical significance of these differences is unclear. Clinicians should be aware that in the presence of leaks, most ICU ventilators require adjustments to maintain an adequate tidal volume.
Similar articles
-
Performance characteristics of seven bilevel mechanical ventilators in pressure-support mode with different cycling criteria: a comparative bench study.Med Sci Monit. 2015 Jan 26;21:310-7. doi: 10.12659/MSM.892080. Med Sci Monit. 2015. PMID: 25619202 Free PMC article.
-
A comparison of leak compensation in acute care ventilators during noninvasive and invasive ventilation: a lung model study.Respir Care. 2013 Dec;58(12):2027-37. doi: 10.4187/respcare.02466. Epub 2013 May 21. Respir Care. 2013. PMID: 23696688
-
Intentional leaks in industrial masks have a significant impact on efficacy of bilevel noninvasive ventilation: a bench test study.Chest. 2009 Mar;135(3):669-677. doi: 10.1378/chest.08-1340. Epub 2008 Oct 10. Chest. 2009. PMID: 18849400
-
Equipment needs for noninvasive mechanical ventilation.Eur Respir J. 2002 Oct;20(4):1029-36. doi: 10.1183/09031936.02.00404202. Eur Respir J. 2002. PMID: 12412700 Review.
-
Are we fully utilizing the functionalities of modern operating room ventilators?Curr Opin Anaesthesiol. 2017 Dec;30(6):698-704. doi: 10.1097/ACO.0000000000000527. Curr Opin Anaesthesiol. 2017. PMID: 28938301 Review.
Cited by
-
Performance characteristics of seven bilevel mechanical ventilators in pressure-support mode with different cycling criteria: a comparative bench study.Med Sci Monit. 2015 Jan 26;21:310-7. doi: 10.12659/MSM.892080. Med Sci Monit. 2015. PMID: 25619202 Free PMC article.
-
Effects of different exhalation valves on CO2 rebreathing and ventilator performance during noninvasive ventilation.Front Med (Lausanne). 2025 Jul 23;12:1538280. doi: 10.3389/fmed.2025.1538280. eCollection 2025. Front Med (Lausanne). 2025. PMID: 40771475 Free PMC article.
-
Scoring System to Evaluate the Performance of ICU Ventilators in the Pandemic of COVID-19: A Lung Model Study.Front Med (Lausanne). 2021 Jul 14;8:663608. doi: 10.3389/fmed.2021.663608. eCollection 2021. Front Med (Lausanne). 2021. PMID: 34336879 Free PMC article.
-
Effects of jet nebulization on ventilator performance with different invasive ventilation modes: A bench study.Front Med (Lausanne). 2022 Oct 19;9:1004551. doi: 10.3389/fmed.2022.1004551. eCollection 2022. Front Med (Lausanne). 2022. PMID: 36341252 Free PMC article.
-
Are all ventilators for NIV performing the same? A bench analysis.J Clin Monit Comput. 2023 Dec;37(6):1497-1511. doi: 10.1007/s10877-023-01019-z. Epub 2023 Jul 31. J Clin Monit Comput. 2023. PMID: 37522978 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials