Comparative Study
. 2009 May;127(5):599-603.
doi: 10.1001/archophthalmol.2009.41.

Comparison of slitlamp optical coherence tomography and scanning peripheral anterior chamber depth analyzer to evaluate angle closure in Asian eyes

Affiliations
Comparative Study

Comparison of slitlamp optical coherence tomography and scanning peripheral anterior chamber depth analyzer to evaluate angle closure in Asian eyes

Hon-Tym Wong et al. Arch Ophthalmol. 2009 May.

Abstract

Objective: To evaluate the effectiveness of slitlamp optical coherence tomography (SL-OCT) and Scanning Peripheral Anterior Chamber depth analyzer (SPAC) in detecting angle closure, using gonioscopy as the reference standard.

Methods: A total of 153 subjects underwent gonioscopy, SL-OCT, and SPAC. The anterior chamber angle (ACA) was classified as closed on gonioscopy if the posterior trabecular meshwork could not be seen; with SL-OCT, closure was determined by contact between the iris and angle wall anterior to the scleral spur; and with SPAC by a numerical grade of 5 or fewer and/or a categorical grade of suspect or potential.

Results: A closed ACA was identified in 51 eyes with gonioscopy, 86 eyes with SL-OCT, and 61 eyes with SPAC (gonioscopy vs SL-OCT, P < .001; gonioscopy vs SPAC, P = .10; SL-OCT vs SPAC, P < .001; McNemar test). Of the 51 eyes with a closed ACA on gonioscopy, SL-OCT detected a closed ACA in 43, whereas SPAC identified 41 (P = .79). An open angle in all 4 quadrants was observed in 102 eyes with gonioscopy, but SL-OCT and SPAC identified 43 and 20 of these eyes, respectively, as having angle closure. The overall sensitivity and specificity for SL-OCT were 84% and 58% vs 80% and 80% for SPAC.

Conclusion: Using gonioscopy as the reference, SL-OCT and SPAC showed good sensitivity for detecting eyes at risk of angle closure.