Intraoperative fluorescence imaging system for on-site assessment of off-pump coronary artery bypass graft
- PMID: 19442948
- DOI: 10.1016/j.jcmg.2008.12.028
Intraoperative fluorescence imaging system for on-site assessment of off-pump coronary artery bypass graft
Abstract
Objectives: The aim of this study was to evaluate the intraoperative fluorescence imaging (IFI) system in the real-time assessment of graft patency during off-pump coronary artery bypass graft.
Background: Intraoperative fluorescence imaging is an intraoperative angiography-like imaging modality using fluorescent indocyanine green excited with laser light. Recently, assessment of graft patency using the IFI system was introduced into clinical use. The feasibility and efficacy of IFI technology in off-pump coronary artery bypass graft has not been systematically compared with other conventional diagnostic modalities.
Methods: Patients undergoing off-pump coronary artery bypass graft received IFI analysis, intraoperative transit time flowmetry, and postoperative X-ray angiography. In off-line IFI analysis, the graft washout was classified based on the number of heartbeats required for indocyanine green washout: fast washout (<or=15 beats) and slow washout (>15 beats).
Results: A total of 507 grafts in 137 patients received IFI analysis. Of all the IFI analyses, 379 (75%) grafts were visualized clearly up to the distal anastomosis. With regard to anastomosis location, anterior location was associated with a higher percentage of fully analyzable images (90%). More than 80% of images were analyzable, irrespective of graft type. Six grafts with acceptable transit time flowmetry results were diagnosed with graft failure by IFI, which required on-site graft revision. All revised grafts' patency was confirmed by post-operative X-ray angiography. Conversely, 21 grafts with unsatisfactory transit time flowmetry results demonstrated acceptable patency with IFI. Graft revision was considered unnecessary in these grafts, and 20 grafts (95%) were patent by post-operative X-ray angiography. Compared with slow washout, fast washout was associated with a higher preoperative ejection fraction, use of internal mammary artery grafts, and anterior anastomosis location.
Conclusions: The IFI system enables on-site assessment of graft patency, providing both morphologic and functional information. This technique may help reduce procedure-related, early graft failures in off-pump bypass patients.
Comment in
-
Assessment of graft patency during coronary artery bypass graft surgery: mitigating the risk.JACC Cardiovasc Imaging. 2009 May;2(5):613-5. doi: 10.1016/j.jcmg.2009.01.009. JACC Cardiovasc Imaging. 2009. PMID: 19442949 No abstract available.
Similar articles
-
Assessment of graft patency during coronary artery bypass graft surgery: mitigating the risk.JACC Cardiovasc Imaging. 2009 May;2(5):613-5. doi: 10.1016/j.jcmg.2009.01.009. JACC Cardiovasc Imaging. 2009. PMID: 19442949 No abstract available.
-
Preliminary experience for the evaluation of the intraoperative graft patency with real color charge-coupled device camera system: an advanced device for simultaneous capturing of color and near-infrared images during coronary artery bypass graft.Interact Cardiovasc Thorac Surg. 2009 Aug;9(2):150-4. doi: 10.1510/icvts.2008.201418. Epub 2009 May 7. Interact Cardiovasc Thorac Surg. 2009. PMID: 19423513
-
A comparison of transit-time flowmetry and intraoperative fluorescence imaging for assessing coronary artery bypass graft patency.J Thorac Cardiovasc Surg. 2005 Aug;130(2):315-20. doi: 10.1016/j.jtcvs.2004.11.033. J Thorac Cardiovasc Surg. 2005. PMID: 16077393
-
Intraoperative angiography leads to graft revision in coronary artery bypass surgery.Ann Thorac Surg. 2004 Aug;78(2):502-5; discussion 505. doi: 10.1016/j.athoracsur.2004.03.004. Ann Thorac Surg. 2004. PMID: 15276506 Review.
-
Techniques and standards in intraoperative graft verification by transit time flow measurement after coronary artery bypass graft surgery: a critical review.Eur J Cardiothorac Surg. 2017 Jan;51(1):26-33. doi: 10.1093/ejcts/ezw203. Epub 2016 Jun 13. Eur J Cardiothorac Surg. 2017. PMID: 27298393 Review.
Cited by
-
Role of Indocyanine Green in Fluorescence Imaging with Near-Infrared Light to Identify Sentinel Lymph Nodes, Lymphatic Vessels and Pathways Prior to Surgery - A Critical Evaluation of Options.Geburtshilfe Frauenheilkd. 2018 Jan;78(1):54-62. doi: 10.1055/s-0043-123937. Epub 2018 Jan 22. Geburtshilfe Frauenheilkd. 2018. PMID: 29375146 Free PMC article.
-
Effectiveness of Intraoperative Indocyanine Green Fluorescence-Navigated Surgery for Superior Mesenteric Vein Thrombosis that Developed During Treatment for Intravascular Lymphoma: A Case Report.Am J Case Rep. 2021 Apr 30;22:e929549. doi: 10.12659/AJCR.929549. Am J Case Rep. 2021. PMID: 33927177 Free PMC article.
-
Ninety-day all-cause emergency room use among coronary artery bypass grafting patients associated with near-infrared fluorescence imaging: a retrospective cohort study.Ann Med Surg (Lond). 2023 Feb 7;85(2):153-160. doi: 10.1097/MS9.0000000000000206. eCollection 2023 Feb. Ann Med Surg (Lond). 2023. PMID: 36845794 Free PMC article.
-
Evaluation of intestinal perfusion by ICG fluorescence imaging in laparoscopic colorectal surgery with DST anastomosis.Surg Endosc. 2017 Mar;31(3):1061-1069. doi: 10.1007/s00464-016-5064-x. Epub 2016 Jun 28. Surg Endosc. 2017. PMID: 27351656
-
Strategies for the coronary surgeon to remain "competitive and co-operative" in the PCI era.Indian Heart J. 2015 Jul-Aug;67(4):351-8. doi: 10.1016/j.ihj.2015.05.002. Epub 2015 Jul 8. Indian Heart J. 2015. PMID: 26304568 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical