Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis
- PMID: 19455003
- DOI: 10.1097/BRS.0b013e3181a4ec2d
Biomechanical effect of constraint in lumbar total disc replacement: a study with finite element analysis
Abstract
Study design: Biomechanical effect of implantation of an artificial disc (AD) was investigated using the nonlinear three-dimensional finite element model of L4-L5. The SB CHARITE and the Prodisc were chosen as the representative prosthesis of unconstrained and constrained ADs (UADs and CADs) and compared with the intact human intervertebral disc.
Objective: To investigate the effect of implantation of an AD to spinal functional unit and to evaluate the difference between the unconstrained and constrained models.
Methods: Intact osteoligamentous L4-L5 finite element model was created with 1-mm computed tomography scan of a cadaveric spine and known material property of each element. Two models implanted with ADs, unconstrained or constrained model, were also developed. The implanted model predictions were compared with that of the intact. Range of motion, force on the spinal ligaments, force on the facet joint, stress on the vertebral body and vertebral endplate with flexion/extension, lateral bending, and axial rotation under 400 N compressive preload were compared among the models.
Results: The implanted models showed increased range of motion in flexion/extension, lateral bending, and axial rotation compared with that of the intact. Under 6-Nm moment, the range of motion were 140%, 170%, and 200% of intact in the UAD model and 133%, 137%, and 138% in the CAD model to each direction of loading. The forces on each ligament were different among the models with various loading conditions. Force on the facet, stress on the vertebral body and vertebral endplate were much larger in implanted model, especially in the CAD model.
Conclusion: By the result of this study it is obvious that implanted segment with AD has large range of motion and suffers from increased loading to surrounding bone and ligaments. The UAD has larger range of motion but exert less loading to the implanted segment than the CAD. It seems that the mobile center of rotation of the UAD has the ability to lessen the facet contact force and stress on the vertebral body.
Similar articles
-
In situ contact analysis of the prosthesis components of Prodisc-L in lumbar spine following total disc replacement.Spine (Phila Pa 1976). 2009 Sep 15;34(20):E716-23. doi: 10.1097/BRS.0b013e3181ae23d1. Spine (Phila Pa 1976). 2009. PMID: 19752690
-
Total disc replacement positioning affects facet contact forces and vertebral body strains.Spine (Phila Pa 1976). 2008 Nov 1;33(23):2510-7. doi: 10.1097/BRS.0b013e318186b258. Spine (Phila Pa 1976). 2008. PMID: 18978591
-
Biomechanical analysis of rotational motions after disc arthroplasty: implications for patients with adult deformities.Spine (Phila Pa 1976). 2006 Sep 1;31(19 Suppl):S152-60. doi: 10.1097/01.brs.0000234782.89031.03. Spine (Phila Pa 1976). 2006. PMID: 16946633
-
Design concepts in lumbar total disc arthroplasty.Eur Spine J. 2008 Dec;17(12):1635-50. doi: 10.1007/s00586-008-0811-x. Epub 2008 Oct 23. Eur Spine J. 2008. PMID: 18946684 Free PMC article. Review.
-
Basic scientific considerations in total disc arthroplasty.Spine J. 2004 Nov-Dec;4(6 Suppl):219S-230S. doi: 10.1016/j.spinee.2004.07.015. Spine J. 2004. PMID: 15541670 Review.
Cited by
-
An 11-year minimum follow-up of the Charite III lumbar disc replacement for the treatment of symptomatic degenerative disc disease.Eur Spine J. 2015 Sep;24(9):2056-64. doi: 10.1007/s00586-015-3939-5. Epub 2015 Apr 21. Eur Spine J. 2015. PMID: 25895882
-
Effects of rod stiffness and fusion mass on the adjacent segments after floating mono-segmental fusion: a study using finite element analysis.Eur Spine J. 2013 May;22(5):1066-77. doi: 10.1007/s00586-012-2611-6. Epub 2012 Dec 16. Eur Spine J. 2013. PMID: 23242620 Free PMC article.
-
A Comparison of the Quality of Sleep between Pre and Post-Surgery Cervical Herniated Nucleus Pulposus Patients Utilizing the Anterior Discectomy Method.Open Access Maced J Med Sci. 2017 Dec 10;5(7):948-954. doi: 10.3889/oamjms.2017.209. eCollection 2017 Dec 15. Open Access Maced J Med Sci. 2017. PMID: 29362625 Free PMC article.
-
Elastic resistance of the spine: Why does motion preservation surgery almost fail?World J Clin Cases. 2013 Jul 16;1(4):134-9. doi: 10.12998/wjcc.v1.i4.134. World J Clin Cases. 2013. PMID: 24303484 Free PMC article.
-
ISASS Policy Statement - Lumbar Artificial Disc.Int J Spine Surg. 2015 Mar 12;9:7. doi: 10.14444/2007. eCollection 2015. Int J Spine Surg. 2015. PMID: 25785243 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous