Corneal flap thickness with the Moria M2 microkeratome and Med-Logics calibrated LASIK blades
- PMID: 19456311
- DOI: 10.1111/j.1755-3768.2008.01500.x
Corneal flap thickness with the Moria M2 microkeratome and Med-Logics calibrated LASIK blades
Abstract
Purpose: This study aimed to compare and study potential factors that affect the accuracy of corneal flap thickness created in laser-assisted in situ keratomileusis (LASIK) using the Moria model 2 (M2) head 130 microkeratome with the Med-Logics calibrated LASIK blades Minus 20 (ML -20) and Minus 30 (ML -30).
Methods: Corneal thickness in 200 (164 myopic and 36 hyperopic) eyes (100 patients) was measured by ultrasonic pachymetry preoperatively and intraoperatively after flap cutting. A total of 100 eyes were treated with the ML -20 and 100 with the ML -30. The right eye was operated before the left eye in each patient, using the same blade. In an additional group of 40 eyes, the left eye was operated first.
Results: Mean corneal flap thickness using the ML -20 blade for an intended flap thickness of 140 μm was 129.1 μm (standard deviation [SD] 15.6, range 104-165 μm) in right eyes and 111.5 μm (SD 14.5, range 78-144 μm) in left eyes. Mean corneal flap thickness using the ML -30 blade for an intended flap thickness of 130 μm was 127.1 μm (SD 16.6, range 90-168 μm) in right eyes and 109.9 μm (SD 16.8, range 72-149 μm) in left eyes.
Conclusions: Both microkeratome blade types cut thinner flaps than were intended. There was substantial variation in flap thickness. The first flap to be cut with a particular blade was considerably thicker than the second flap cut with the same blade. Based on these data, we recommend the use of disposable single-use microkeratomes rather than these ML blades.
Similar articles
-
Corneal flap thickness with the Moria M2 single-use head 90 microkeratome.Acta Ophthalmol Scand. 2007 Jun;85(4):401-6. doi: 10.1111/j.1600-0420.2006.00838.x. Acta Ophthalmol Scand. 2007. PMID: 17559465
-
Microkeratome blades and corneal flap thickness in LASIK.Ophthalmic Surg Lasers Imaging. 2008 Nov-Dec;39(6):471-5. doi: 10.3928/15428877-20081101-19. Ophthalmic Surg Lasers Imaging. 2008. PMID: 19065977
-
Comparison of corneal flap thickness between primary eyes and fellow eyes using the Zyoptix XP microkeratome.J Cataract Refract Surg. 2007 Dec;33(12):2049-53. doi: 10.1016/j.jcrs.2007.07.027. J Cataract Refract Surg. 2007. PMID: 18053902
-
Corneal flap thickness in laser in situ keratomileusis using the Moria M2 microkeratome.J Cataract Refract Surg. 2004 Sep;30(9):1902-8. doi: 10.1016/j.jcrs.2004.01.017. J Cataract Refract Surg. 2004. PMID: 15342053
-
[Advantages and disadvantages of the femtosecond laser microkeratome].J Fr Ophtalmol. 2005 May;28(5):535-46. doi: 10.1016/s0181-5512(05)81094-0. J Fr Ophtalmol. 2005. PMID: 15976724 Review. French.
Cited by
-
Temporal and Spatial Flap Variability in Laser In-Situ Keratomileusis by Optical Coherence Tomography.J Ophthalmic Vis Res. 2017 Oct-Dec;12(4):368-373. doi: 10.4103/jovr.jovr_173_16. J Ophthalmic Vis Res. 2017. PMID: 29090044 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources