In vivo repeatability of automated volume calculations of small pulmonary nodules with CT
- PMID: 19457831
- DOI: 10.2214/AJR.08.1825
In vivo repeatability of automated volume calculations of small pulmonary nodules with CT
Abstract
Objective: The objectives of our study were to evaluate the in vivo reproducibility of automated volume calculations of small lung nodules with both low-dose and standard-dose CT and to assess whether repeatability within each technique varies according to the diameter, site, or morphology of the nodule or to percentage of emphysema.
Subjects and methods: Sixty-six subjects with 83 solid pulmonary nodules between 5 and 10 mm in diameter were enrolled in this prospective study. Four consecutive MDCT data sets, two low dose and two standard dose, were obtained for each nodule on separate breath-holds during the same session. The volume of each nodule was calculated by automated software. Repeatability was evaluated by Bland-Altman's approach and the coefficient of repeatability. Associations of the percentage of volume variation between two measurements with nodule diameter, emphysema percentage, nodule site, and nodule morphology were assessed by Spearman's correlation coefficient and the Kruskal-Wallis test. A p value of < 0.05 was considered statistically significant.
Results: The range of variation of the volumes of pulmonary nodules between two subsequent measurements was -38% +/- 60% for low-dose CT and -27% +/- 40% for standard-dose CT. No significant statistical association was found between variation in volume measurements and nodule site, nodule diameter, nodule morphology, or emphysema percentage by semiautomated calculation of lung density.
Conclusion: Automated volume calculations of small pulmonary nodules can significantly differ between two subsequent breath-holds with both low-dose and standard-dose CT techniques; in clinical practice we recommend that a volume variation of greater than 30% for nodules between 5 and 10 mm should be confirmed by follow-up CT to be sure that a nodule is actually growing.
Similar articles
-
Pulmonary nodule volumetric measurement variability as a function of CT slice thickness and nodule morphology.AJR Am J Roentgenol. 2007 Feb;188(2):306-12. doi: 10.2214/AJR.05.1063. AJR Am J Roentgenol. 2007. PMID: 17242235
-
Pulmonary nodules: Preliminary experience with semiautomated volumetric evaluation by CT stratum.Acad Radiol. 2010 Jul;17(7):900-11. doi: 10.1016/j.acra.2010.03.003. Epub 2010 May 5. Acad Radiol. 2010. PMID: 20447841
-
Semi-automatic volumetric measurement of lung cancer using multi-detector CT effects of nodule characteristics.Acad Radiol. 2009 Oct;16(10):1179-86. doi: 10.1016/j.acra.2009.04.007. Epub 2009 Jun 12. Acad Radiol. 2009. PMID: 19524456
-
Noncalcified lung nodules: volumetric assessment with thoracic CT.Radiology. 2009 Apr;251(1):26-37. doi: 10.1148/radiol.2511071897. Radiology. 2009. PMID: 19332844 Free PMC article. Review.
-
Multidetector CT of solitary pulmonary nodules.Radiol Clin North Am. 2010 Jan;48(1):141-55. doi: 10.1016/j.rcl.2009.09.005. Radiol Clin North Am. 2010. PMID: 19995633 Review.
Cited by
-
Pulmonary Nodules: growth rate assessment in patients by using serial CT and three-dimensional volumetry.Radiology. 2012 Feb;262(2):662-71. doi: 10.1148/radiol.11100878. Epub 2011 Dec 9. Radiology. 2012. PMID: 22156993 Free PMC article.
-
Variability in CT lung-nodule volumetry: Effects of dose reduction and reconstruction methods.Med Phys. 2015 May;42(5):2679-89. doi: 10.1118/1.4918919. Med Phys. 2015. PMID: 25979066 Free PMC article.
-
Deep-learning reconstruction for ultra-low-dose lung CT: Volumetric measurement accuracy and reproducibility of artificial ground-glass nodules in a phantom study.Br J Radiol. 2022 Feb 1;95(1130):20210915. doi: 10.1259/bjr.20210915. Epub 2021 Dec 15. Br J Radiol. 2022. PMID: 34908478 Free PMC article.
-
Lung nodules: size still matters.Eur Respir Rev. 2017 Dec 20;26(146):170025. doi: 10.1183/16000617.0025-2017. Print 2017 Dec 31. Eur Respir Rev. 2017. PMID: 29263171 Free PMC article. Review.
-
Inter- and intrascanner variability of pulmonary nodule volumetry on low-dose 64-row CT: an anthropomorphic phantom study.Br J Radiol. 2013 Sep;86(1029):20130160. doi: 10.1259/bjr.20130160. Epub 2013 Jul 24. Br J Radiol. 2013. PMID: 23884758 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Medical