[Serial tomotherapy vs. MLC-IMRT (multileaf collimator intensity modulated radiotherapy) for simultaneous boost treatment large intracerebral lesions]
- PMID: 19459586
- DOI: 10.1016/j.zemedi.2008.07.003
[Serial tomotherapy vs. MLC-IMRT (multileaf collimator intensity modulated radiotherapy) for simultaneous boost treatment large intracerebral lesions]
Abstract
Introduction: Recent data suggest that a radiosurgery boost treatment for up to three brain metastases in addition to whole brain radiotherapy (WBRT) is beneficial. Sequential treatment of multiple metastatic lesions is time-consuming and optimal normal tissue sparing is not trivial for larger metastases when separate plans are created and are only superimposed afterwards. Sequential Tomotherapy (see image I) with noncoplanar arcs and Multi-field IMRT may streamline the process and enable easy simultaneous treatment. We compared plans for 2-3 intracerebral targets calculated with Intensity Modulated Radiotherapy (IMRT) based on treatment with MLC or sequential Tomotherapy using the Peacock-System (see image II). Treatment time was not to exceed 90 min on a linac with standard dose rate. MIMiC plans without treatment-time restrictions were created as a benchmark.
Materials and methods: Calculations are based on a Siemens KD2 linac with a dose rate of 200 MU/min. Step-and-Shoot IMRT is performed with a standard MLC (2 x 29 leaves, 1 cm), serial Tomotherapy with the Multivane-Collimator MIMiC (NOMOS Inc. USA) (see image II). Treatment plans are created with Corvus 5.0. To create plans with good conformity we chose a noncoplanar beam- and arc geometry for each approach (IMRT 4-, MIMiC 5-couch angles). The benchmark MIMiC plans with maximally steep dose gradients had 9 couch angles. For plan comparison reasons, 10 Gy were prescribed to 90% of the PTV. Steepness of dose gradients, homogeneity and conformity were assessed by the following parameters: Volume encompassed by certain isodoses outside the target as well as homogeneity and conformity as indicated by Homogeneity- and Conformity-Index.
Results: Plans without treatment-time restrictions had slightest dose to organ at risk (OAR), normal tissue and least Conformity-index. MIMiC- and MLC-IMRT based plans can be treated within the intended period of 90 min, all plans met the required dose (see Table 2). MLC based plans resulted in higher dose to organs at risk (OAR) (see table 1) and dose to tissue outside the targets (see table 3), as indicated by a higher CI (see image III). The HI was similar for all calculated plans (see image IV).
Discussion: When treatment plans resulting in a similar treatment time were compared, serial Tomotherapy showed minor advantages over MLC based IMRT with regard to conformity, OAR sparing, and steepness of dose gradients. Both methods are inferior to serial Tomotherapy with ideal plan quality disregarding treatment efficiency. Treating multiple metastases in less than 1 h would therefore be possible on a LINAC with high dose rate and bidirectional rotation with minor compromises on gradient steepness.
Similar articles
-
Evaluation of volumetric modulated arc therapy for cranial radiosurgery using multiple noncoplanar arcs.Med Phys. 2011 Nov;38(11):5863-72. doi: 10.1118/1.3641874. Med Phys. 2011. PMID: 22047350
-
Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer.Radiother Oncol. 2009 Nov;93(2):226-33. doi: 10.1016/j.radonc.2009.08.011. Epub 2009 Sep 16. Radiother Oncol. 2009. PMID: 19765846
-
Treatment planning strategy for whole-brain radiotherapy with hippocampal sparing and simultaneous integrated boost for multiple brain metastases using intensity-modulated arc therapy.Med Dosim. 2016 Winter;41(4):315-322. doi: 10.1016/j.meddos.2016.08.001. Epub 2016 Sep 28. Med Dosim. 2016. PMID: 27692518
-
Helical tomotherapy for whole-brain irradiation with integrated boost to multiple brain metastases: evaluation of dose distribution characteristics and comparison with alternative techniques.Int J Radiat Oncol Biol Phys. 2013 Jul 15;86(4):734-42. doi: 10.1016/j.ijrobp.2013.03.031. Epub 2013 May 14. Int J Radiat Oncol Biol Phys. 2013. PMID: 23680034 Clinical Trial.
-
Clinical implementation of adaptive helical tomotherapy: a unique approach to image-guided intensity modulated radiotherapy.Technol Cancer Res Treat. 2006 Oct;5(5):465-79. doi: 10.1177/153303460600500503. Technol Cancer Res Treat. 2006. PMID: 16981789 Review.
Cited by
-
Intensity-modulated radiation therapy combined with concomitant temozolomide for brain metastases from lung adenocarcinoma.Oncol Lett. 2018 Oct;16(4):4285-4290. doi: 10.3892/ol.2018.9171. Epub 2018 Jul 18. Oncol Lett. 2018. PMID: 30214562 Free PMC article.
-
On the cutting edge of intensity modulated radiotherapy and simultaneous integrated boost (IMRT-SIB): The case of a patient with 8 brain metastases.Rep Pract Oncol Radiother. 2015 Jul-Aug;20(4):316-9. doi: 10.1016/j.rpor.2014.09.001. Epub 2014 Sep 23. Rep Pract Oncol Radiother. 2015. PMID: 26109921 Free PMC article.
-
Use of Helical TomoTherapy for the Focal Hypofractionated Treatment of Limited Brain Metastases in the Initial and Recurrent Setting.Front Oncol. 2015 Feb 9;5:27. doi: 10.3389/fonc.2015.00027. eCollection 2015. Front Oncol. 2015. PMID: 25709970 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Medical
Research Materials
Miscellaneous