Abstract
The objective of this paper is to validate angle independent vector velocity methods for blood velocity estimation. Conventional Doppler ultrasound (US) only estimates the blood velocity along the US beam direction where the estimate is angle corrected assuming laminar flow parallel to vessel boundaries. This results in incorrect blood velocity estimates, when angle of insonation approaches 90 degrees or when blood flow is non-laminar. Three angle independent vector velocity methods are evaluated in this paper: directional beamforming (DB), synthetic aperture flow imaging (STA) and transverse oscillation (TO). The performances of the three methods were investigated by measuring the stroke volume in the right common carotid artery of 11 healthy volunteers with magnetic resonance phase contrast angiography (MRA) as reference. The correlation with confidence intervals (CI) between the three vector velocity methods and MRA were: DB vs. MRA: R=0.84 (p<0.01, 95% CI: 0.49-0.96); STA vs. MRA: R=0.71 (p<0.05, 95% CI: 0.19-0.92) and TO vs. MRA: R=0.91 (p<0.01, 95% CI: 0.69-0.98). No significant differences were observed for any of the three comparisons (DB vs. MRA: p=0.65; STA vs. MRA: p=0.24; TO vs. MRA: p=0.36). Bland-Altman plots were additionally constructed, and mean differences with limits of agreements (LoA) for the three comparisons were: DB vs. MRA=0.17 ml (95% CI: -0.61-0.95) with LoA=-2.11-2.44 ml; STA vs. MRA=-0.55 ml (95% CI: -1.54-0.43) with LoA=-3.42-2.32 ml; TO vs. MRA=0.24 ml (95% CI: -0.32-0.81) with LoA=-1.41-1.90 ml. According to the results, reliable volume flow estimates can be obtained with all three methods. The three US vector velocity techniques can yield quantitative insight into flow dynamics and visualize complex flow patterns, which potentially can give the clinician a novel tool for cardiovascular disease assessment.
Similar articles
-
In vivo validation of a blood vector velocity estimator with MR angiography.IEEE Trans Ultrason Ferroelectr Freq Control. 2009 Jan;56(1):91-100. doi: 10.1109/TUFFC.2009.1008. IEEE Trans Ultrason Ferroelectr Freq Control. 2009. PMID: 19213635
-
A Comparison Study of Vector Velocity, Spectral Doppler and Magnetic Resonance of Blood Flow in the Common Carotid Artery.Ultrasound Med Biol. 2018 Aug;44(8):1751-1761. doi: 10.1016/j.ultrasmedbio.2018.05.002. Epub 2018 May 24. Ultrasound Med Biol. 2018. PMID: 29804906
-
Accuracy and reproducibility of a novel dual-beam vector Doppler method.Ultrasound Med Biol. 2009 May;35(5):829-38. doi: 10.1016/j.ultrasmedbio.2008.10.012. Epub 2008 Dec 24. Ultrasound Med Biol. 2009. PMID: 19110369
-
Non-contrast enhanced MR angiography: physical principles.J Magn Reson Imaging. 2012 Aug;36(2):286-304. doi: 10.1002/jmri.23641. J Magn Reson Imaging. 2012. PMID: 22807222 Review.
-
Three-Dimensional Carotid Plaque MR Imaging.Neuroimaging Clin N Am. 2016 Feb;26(1):1-12. doi: 10.1016/j.nic.2015.09.001. Epub 2015 Oct 19. Neuroimaging Clin N Am. 2016. PMID: 26610656 Free PMC article. Review.
Cited by
-
Quantitative Magnetic Resonance Angiography in Internal Carotid Artery Occlusion with Primary Collateral Pathway.J Stroke. 2015 Sep;17(3):320-6. doi: 10.5853/jos.2015.17.3.320. Epub 2015 Sep 30. J Stroke. 2015. PMID: 26437997 Free PMC article.
-
Vector velocity estimation of blood flow - A new application in medical ultrasound.Ultrasound. 2017 Nov;25(4):189-199. doi: 10.1177/1742271X17713353. Epub 2017 Jun 5. Ultrasound. 2017. PMID: 29163655 Free PMC article. Review.
-
Recent developments in vascular ultrasound technology.Ultrasound. 2015 Aug;23(3):158-65. doi: 10.1177/1742271X15578778. Epub 2015 Mar 26. Ultrasound. 2015. PMID: 27433252 Free PMC article. Review.
-
Common Carotid Artery Volume Flow: A Comparison Study between Ultrasound Vector Flow Imaging and Phase Contrast Magnetic Resonance Imaging.Neurol Int. 2021 Jun 23;13(3):269-278. doi: 10.3390/neurolint13030028. Neurol Int. 2021. PMID: 34201493 Free PMC article.
-
Ultrasound Open Platforms for Next-Generation Imaging Technique Development.IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Jul;65(7):1078-1092. doi: 10.1109/TUFFC.2018.2844560. IEEE Trans Ultrason Ferroelectr Freq Control. 2018. PMID: 29993364 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous