Bone remodeling around porous metal cementless acetabular components
- PMID: 19473807
- DOI: 10.1016/j.arth.2009.04.025
Bone remodeling around porous metal cementless acetabular components
Abstract
Bone remodeling around cementless acetabular components after total hip arthroplasty has not been well characterized. A randomized, prospective study of total hip arthroplasty was performed comparing 2 cementless acetabular implants: a solid titanium and a more elastic porous tantalum design. Seventeen hips (9 porous tantalum, 8 titanium) underwent quantitative computed tomography at mean of 7.7 years, and adjacent bone mineral density (BMD) was calculated. The absolute and relative decrease in BMD from preoperative level was less in zones 9 to 15 mm adjacent to the porous tantalum compared to the titanium component (P <or= .02) and predominated posterosuperiorly. The relative BMD increased in all regions adjacent to the porous tantalum component from 5% to 40% over the control. This data demonstrates stress-shielding likely occurs less around a highly porous metal implant of material with an elastic modulus similar to bone.
Copyright (c) 2010 Elsevier Inc. All rights reserved.
Similar articles
-
Periprosthetic mineralization around cementless total hip endoprosthesis: longitudinal study and cross-sectional study on titanium threaded acetabular cup and cementless Spotorno stem with DEXA.Calcif Tissue Int. 1998 Feb;62(2):177-82. doi: 10.1007/s002239900413. Calcif Tissue Int. 1998. PMID: 9437053
-
Radiological evaluation of the metal-bone interface of a porous tantalum monoblock acetabular component.J Bone Joint Surg Br. 2006 Mar;88(3):304-9. doi: 10.1302/0301-620X.88B3.16940. J Bone Joint Surg Br. 2006. PMID: 16498001
-
Performance of Porous Tantalum vs. Titanium Cup in Total Hip Arthroplasty: Randomized Trial with Minimum 10-Year Follow-Up.J Arthroplasty. 2015 Jun;30(6):1008-13. doi: 10.1016/j.arth.2015.01.013. Epub 2015 Jan 21. J Arthroplasty. 2015. PMID: 25765132 Clinical Trial.
-
Use of porous tantalum for acetabular reconstruction in revision hip arthroplasty.J Bone Joint Surg Am. 2013 Nov 6;95(21):1981-7. doi: 10.2106/JBJS.L.01313. J Bone Joint Surg Am. 2013. PMID: 24196469 Review.
-
Change in CT-measured acetabular bone density following total hip arthroplasty: a systematic review and meta-analysis.Acta Orthop. 2023 Apr 27;94:191-199. doi: 10.2340/17453674.2023.11635. Acta Orthop. 2023. PMID: 37114317 Free PMC article.
Cited by
-
Total Hip Arthroplasty Using Modular Trabecular Metal Acetabular Components for Failed Treatment of Acetabular Fractures: A Mid-term Follow-up Study.Chin Med J (Engl). 2016 Apr 20;129(8):903-8. doi: 10.4103/0366-6999.179793. Chin Med J (Engl). 2016. PMID: 27064033 Free PMC article.
-
An evaluation of the potential consequences of drilling titanium and tantalum implants during surgery - a pilot study.BMC Musculoskelet Disord. 2017 Nov 3;18(1):426. doi: 10.1186/s12891-017-1784-x. BMC Musculoskelet Disord. 2017. Retraction in: BMC Musculoskelet Disord. 2019 Oct 24;20(1):469. doi: 10.1186/s12891-019-2824-5. PMID: 29100511 Free PMC article. Retracted.
-
Radiographic and clinical outcomes of second generation Trabecular Metal™ glenoid components in total shoulder arthroplasty.Shoulder Elbow. 2021 Aug;13(4):416-425. doi: 10.1177/1758573220909981. Epub 2020 Mar 24. Shoulder Elbow. 2021. PMID: 34394739 Free PMC article.
-
Additive manufacturing technique-designed metallic porous implants for clinical application in orthopedics.RSC Adv. 2018 Jul 16;8(44):25210-25227. doi: 10.1039/c8ra04815k. eCollection 2018 Jul 9. RSC Adv. 2018. PMID: 35542139 Free PMC article. Review.
-
Uncemented Cups and Impaction Bone Grafting for Acetabular Bone Loss in Revision Hip Arthroplasty: A Review of Rationale, Indications, and Outcomes.Materials (Basel). 2022 May 23;15(10):3728. doi: 10.3390/ma15103728. Materials (Basel). 2022. PMID: 35629756 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical