Abstract
We carried out metal artefact-reduction MRI, three-dimensional CT measurement of the position of the component and inductively-coupled plasma mass spectrometry analysis of cobalt and chromium levels in whole blood on 26 patients with unexplained pain following metal-on-metal resurfacing arthroplasty. MRI showed periprosthetic lesions around 16 hips, with 14 collections of fluid and two soft-tissue masses. The lesions were seen in both men and women and in symptomatic and asymptomatic hips. Using three-dimensional CT, the median inclination of the acetabular component was found to be 55 degrees and its positioning was outside the Lewinnek safe zone in 13 of 16 cases. Using inductively-coupled plasma mass spectrometry, the levels of blood metal ions tended to be higher in painful compared with well-functioning metal-on-metal hips. These three clinically useful investigations can help to determine the cause of failure of the implant, predict the need for future revision and aid the choice of revision prostheses.
Similar articles
-
Are Females at Greater Risk for Revision Surgery After Hip Resurfacing Arthroplasty With the Articular Surface Replacement Prosthesis?Clin Orthop Relat Res. 2016 Oct;474(10):2257-65. doi: 10.1007/s11999-016-4860-x. Epub 2016 Apr 27. Clin Orthop Relat Res. 2016. PMID: 27121872 Free PMC article.
-
Understanding why metal-on-metal hip arthroplasties fail: a comparison between patients with well-functioning and revised birmingham hip resurfacing arthroplasties. AAOS exhibit selection.J Bone Joint Surg Am. 2012 Feb 15;94(4):e22. doi: 10.2106/JBJS.K.01266. J Bone Joint Surg Am. 2012. PMID: 22336981
-
Revision of Metal-on-metal Hip Prostheses Results in Marked Reduction of Blood Cobalt and Chromium Ion Concentrations.Clin Orthop Relat Res. 2015 Jul;473(7):2305-13. doi: 10.1007/s11999-015-4156-6. Epub 2015 Jan 27. Clin Orthop Relat Res. 2015. PMID: 25623595 Free PMC article.
-
Chromium and cobalt ion release following the Durom high carbon content, forged metal-on-metal surface replacement of the hip.J Bone Joint Surg Br. 2007 Apr;89(4):441-8. doi: 10.1302/0301-620X.89B4.18054. J Bone Joint Surg Br. 2007. PMID: 17463109 Review.
-
Systemic toxicity related to metal hip prostheses.Clin Toxicol (Phila). 2014 Sep-Oct;52(8):837-47. doi: 10.3109/15563650.2014.944977. Epub 2014 Aug 16. Clin Toxicol (Phila). 2014. PMID: 25132471 Review.
Cited by
-
Grading the severity of soft tissue changes associated with metal-on-metal hip replacements: reliability of an MR grading system.Skeletal Radiol. 2011 Mar;40(3):303-7. doi: 10.1007/s00256-010-1000-7. Epub 2010 Jul 25. Skeletal Radiol. 2011. PMID: 20658133 Clinical Trial.
-
Imaging near orthopedic hardware.J Magn Reson Imaging. 2017 Jul;46(1):24-39. doi: 10.1002/jmri.25577. Epub 2017 Feb 2. J Magn Reson Imaging. 2017. PMID: 28152257 Free PMC article. Review.
-
What is appropriate surveillance for metal-on-metal hip arthroplasty patients?Acta Orthop. 2018 Feb;89(1):29-39. doi: 10.1080/17453674.2017.1398011. Epub 2017 Nov 6. Acta Orthop. 2018. PMID: 29105547 Free PMC article.
-
The effects on bone cells of metal ions released from orthopaedic implants. A review.Clin Cases Miner Bone Metab. 2013 Jan;10(1):34-40. doi: 10.11138/ccmbm/2013.10.1.034. Clin Cases Miner Bone Metab. 2013. PMID: 23858309 Free PMC article.
-
Do ion levels in hip resurfacing differ from metal-on-metal THA at midterm?Clin Orthop Relat Res. 2011 Jan;469(1):180-7. doi: 10.1007/s11999-010-1424-3. Epub 2010 Jun 11. Clin Orthop Relat Res. 2011. PMID: 20544315 Free PMC article.
MeSH terms
Substances
LinkOut - more resources