A simple and effective method for validation and measurement of collimator output factors for Leksell Gamma Knife Perfexion
- PMID: 19491451
- DOI: 10.1088/0031-9155/54/12/019
A simple and effective method for validation and measurement of collimator output factors for Leksell Gamma Knife Perfexion
Abstract
Accurate determination of collimator output factors is important for Leksell Gamma Knife radiosurgery. The new Leksell Gamma Knife Perfexion system has a completely redesigned collimator system and the collimator output factors are different from the other Leksell Gamma Knife models. In this study, a simple method was developed to validate the collimator output factors specifically for Leksell Gamma Knife Perfexion. The method uses double-shot exposures on a single film to eliminate repeated setups and the necessity to acquire dose calibration curves required for the traditional film exposure method. Using the method, the collimator output factors with respect to the 16 mm collimator were measured to be 0.929 +/- 0.009 and 0.817 +/- 0.012 for the 8 mm and the 4 mm collimator, respectively. These values are in agreement (within 2%) with the default values of 0.924 and 0.805 in the Leksell Gamma Plan treatment planning system. These values also agree with recently published results of 0.917 (8 mm collimator) and 0.818 (4 mm collimator) obtained from the traditional methods. Given the efficiency of the method, measurement and validation of the collimator output factors can be readily adopted in commissioning and quality assurance of a Leksell Gamma Knife Perfexion system.
Similar articles
-
The Leksell Gamma Knife Perfexion and comparisons with its predecessors.Neurosurgery. 2007 Sep;61(3 Suppl):130-40; discussion 140-1. doi: 10.1227/01.neu.0000289726.35330.8a. Neurosurgery. 2007. PMID: 17876243
-
Determination of output factors for the Leksell gamma knife using ion chamber, thermoluminescence detectors and films.J BUON. 2006 Apr-Jun;11(2):223-7. J BUON. 2006. PMID: 17318975
-
Verification of source and collimator configuration for Gamma Knife Perfexion using panoramic imaging.Med Phys. 2010 Mar;37(3):1325-31. doi: 10.1118/1.3327458. Med Phys. 2010. PMID: 20384269
-
3-D treatment planning system-Leksell Gamma Knife treatment planning system.Med Dosim. 2018 Summer;43(2):177-183. doi: 10.1016/j.meddos.2018.03.001. Epub 2018 Apr 12. Med Dosim. 2018. PMID: 29656889 Review.
-
The gamma knife in ophthalmology. Part One--Uveal melanoma.Klin Oczna. 2014;116(2):130-4. Klin Oczna. 2014. PMID: 25345293 Review.
Cited by
-
Patient-specific independent 3D GammaPlan quality assurance for Gamma Knife Perfexion radiosurgery.J Appl Clin Med Phys. 2013 Jan 7;14(1):3949. doi: 10.1120/jacmp.v14i1.3949. J Appl Clin Med Phys. 2013. PMID: 23318382 Free PMC article.
-
PRESAGE 3D dosimetry accurately measures Gamma Knife output factors.Phys Med Biol. 2014 Dec 7;59(23):N211-20. doi: 10.1088/0031-9155/59/23/N211. Epub 2014 Nov 4. Phys Med Biol. 2014. PMID: 25368961 Free PMC article.
-
A novel and effective method for validation and measurement of output factors for Leksell Gamma Knife® Icon™ using TRS 483 protocol.J Appl Clin Med Phys. 2020 Oct;21(10):80-88. doi: 10.1002/acm2.13011. Epub 2020 Sep 6. J Appl Clin Med Phys. 2020. PMID: 32892452 Free PMC article.
-
Characterization and validation of an intra-fraction motion management system for masked-based radiosurgery.J Appl Clin Med Phys. 2019 May;20(5):21-26. doi: 10.1002/acm2.12573. Epub 2019 May 4. J Appl Clin Med Phys. 2019. PMID: 31055877 Free PMC article.
-
Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.Neuro Oncol. 2017 Apr 1;19(suppl_2):ii2-ii15. doi: 10.1093/neuonc/nox001. Neuro Oncol. 2017. PMID: 28380635 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources