Correction of myopia after cataract surgery with a light-adjustable lens
- PMID: 19500852
- DOI: 10.1016/j.ophtha.2009.02.012
Correction of myopia after cataract surgery with a light-adjustable lens
Abstract
Purpose: To determine whether residual myopia could be corrected postoperatively using the light-adjustable lens (LAL) technology in patients undergoing cataract surgery and LAL implantation.
Design: A prospective clinical study was conducted at Codet Vision Institute in Tijuana, Mexico. The LALs were implanted that would purposely result in myopic errors of up to -1.5 D (diopter). The LAL was treated with a spatial intensity profile delivered by a digital light delivery device to induce a targeted myopic refractive change. Once the desirable myopic correction was achieved, the LAL was treated again to lock-in the lens power.
Participants: Fourteen eyes of 14 patients were studied.
Methods: The manifest refraction, uncorrected visual acuity (UCVA), and best- or spectacle-corrected visual acuity (BCVA), were measured with follow up time of 1 to 9 months to determine the achieved refractive corrections and their stability.
Main outcome measures: We measured UCVA and BCVA, achieved versus targeted refractive outcome, and refractive stability with follow up time of 1 to 9 months.
Results: Of 14 eyes, 13 eyes (92.9%) achieved 0.25 D of the target refraction at 1 day post lock-in with 100% of the eyes achieving the targeted refractive adjustment within 0.5 D or better with up to 9 months postoperative follow-up. All eyes treated show no change in manifest spherical refraction >0.25 D between 1 day post lock-in, and 3, 6, and 9 months postoperative visits. The data demonstrate the stability of the achieved refractive change after the adjustment and lock-in procedures. The mean rate of change was 0.006 D per month, which is 6 times more stable than that of laser corneal refractive procedures.
Conclusions: Residual myopia errors up to -1.5 D were successfully corrected with precision and significant improvement in UCVA and without compromising BCVA using the LAL technology.
Similar articles
-
Correction of residual hyperopia after cataract surgery using the light adjustable intraocular lens technology.Am J Ophthalmol. 2009 Mar;147(3):392-397.e1. doi: 10.1016/j.ajo.2008.08.039. Epub 2008 Nov 20. Am J Ophthalmol. 2009. PMID: 19019340 Clinical Trial.
-
Clinical evaluation of an ultraviolet light adjustable intraocular lens implanted after cataract removal: eighteen months follow-up.Ophthalmology. 2011 Dec;118(12):2382-8. doi: 10.1016/j.ophtha.2011.05.030. Epub 2011 Aug 27. Ophthalmology. 2011. PMID: 21872939
-
Refractive lens exchange with foldable toric intraocular lens.Am J Ophthalmol. 2009 Jun;147(6):990-6, 996.e1. doi: 10.1016/j.ajo.2009.01.004. Epub 2009 Mar 14. Am J Ophthalmol. 2009. PMID: 19286153 Clinical Trial.
-
Phakic intraocular lens implantation for the correction of myopia: a report by the American Academy of Ophthalmology.Ophthalmology. 2009 Nov;116(11):2244-58. doi: 10.1016/j.ophtha.2009.08.018. Ophthalmology. 2009. PMID: 19883852 Review.
-
Refractive lens exchange combined with pars plana vitrectomy to correct high myopia.Eye (Lond). 2006 Jun;20(6):655-60. doi: 10.1038/sj.eye.6701933. Epub 2005 Jun 10. Eye (Lond). 2006. PMID: 15951756 Review.
Cited by
-
Corrective Techniques and Future Directions for Treatment of Residual Refractive Error Following Cataract Surgery.Expert Rev Ophthalmol. 2014 Dec;9(6):529-537. doi: 10.1586/17469899.2014.966817. Expert Rev Ophthalmol. 2014. PMID: 25663845 Free PMC article.
-
Astigmatic correction with implantation of a light adjustable vs monofocal lens: a single site analysis of a randomized controlled trial.Int J Ophthalmol. 2019 Jul 18;12(7):1101-1107. doi: 10.18240/ijo.2019.07.08. eCollection 2019. Int J Ophthalmol. 2019. PMID: 31341799 Free PMC article.
-
Evaluating the evidence for and against the use of IOLs in infants and young children.Expert Rev Med Devices. 2016;13(4):381-9. doi: 10.1586/17434440.2016.1153967. Epub 2016 Feb 29. Expert Rev Med Devices. 2016. PMID: 26878234 Free PMC article. Review.
-
Correlation between contrast sensitivity and higher-order aberration based on pupil diameter after cataract surgery.Clin Ophthalmol. 2011;5:1701-7. doi: 10.2147/OPTH.S21819. Epub 2011 Dec 7. Clin Ophthalmol. 2011. PMID: 22205829 Free PMC article.
-
Visual Outcomes of an Enhanced UV Protected Light Adjustable Lens Using a Novel Co-Managed, Open-Access Methodology.Clin Ophthalmol. 2022 Aug 4;16:2413-2420. doi: 10.2147/OPTH.S378525. eCollection 2022. Clin Ophthalmol. 2022. PMID: 35957659 Free PMC article.
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical