Comparative Study
. 2009 Aug;74(2):427-30.
doi: 10.1016/j.urology.2009.01.007. Epub 2009 Jun 6.

Prospective comparison of flexible fiberoptic and digital cystoscopes

Affiliations
Comparative Study

Prospective comparison of flexible fiberoptic and digital cystoscopes

Zhamshid Okhunov et al. Urology. 2009 Aug.

Abstract

Objectives: To compare the optics, performance, and durability of digital (DC) and fiberoptic (FC) cystoscopes.

Methods: In an office setting, we randomly assigned staff urologists to 1 of the 4 cystoscopes, the Gyrus-ACMI ICN-0564 (AD), Gyrus-ACMI ACN-2 (AF), Olympus CYF-V2 EndoEYE Cysto-Nephro Videoscope (OD), Olympus CYF-5 Cysto-Fiberscope (OF), to perform diagnostic or surveillance cystoscopy and stent removal. The documented metrics included a subjective surgeon assessment of cystoscope optics and function characteristics on a 10-point scale (1, poor to 10, excellent). The measurement of the upward and downward cystoscope deflection and damage and repairs were all documented.

Results: A total of 1022 cases were performed. The DC and FC were used 690 and 332 times, respectively. Two repairs (0.2%) were documented (1 AF and 1 AD); both resulted directly from incorrect cystoscope handling/cleaning. The mean operative time per case for the DC and FC was 4.5 and 4.6 minutes, respectively (P = .66). The mean surgeon optical ranking for the DC and FC was 8.4 and 7.8, respectively (P = .0076). The mean surgeon deflection ranking for the DC and FC was 8.6 and 8.0, respectively (P = .0001). The mean surgeon retroflex deflection ranking for the DC and FC was 8.4 and 7.8, respectively (P = .001). The mean overall cystoscope score surgeon ranking for the DC and FC was 8.6 and 7.9, respectively (P = .0001).

Conclusions: In the office setting, with proper care, FCs and DCs are durable for office applications. Overall, surgeons significantly preferred the DCs as demonstrated by discrepancies in both use and differences in the subjective metrics.

PubMed Disclaimer