Randomized Controlled Trial
. 2009 Jul;111(1):97-101.
doi: 10.1097/ALN.0b013e3181a8649f.

Cervical spine motion: a fluoroscopic comparison of the AirTraq Laryngoscope versus the Macintosh laryngoscope

Affiliations
Randomized Controlled Trial

Cervical spine motion: a fluoroscopic comparison of the AirTraq Laryngoscope versus the Macintosh laryngoscope

Timothy P Turkstra et al. Anesthesiology. 2009 Jul.

Abstract

Background: The optimal technique to intubate the trachea in patients presenting with a potential or documented cervical spine (C-spine) injury remains unresolved. Using continuous fluoroscopic video assessment, C-spine motion during laryngoscopy with an AirTraq Laryngoscope (King Medical Systems, Newark, DE) was compared to that with intubation using a Macintosh blade.

Methods: Twenty-four healthy surgical patients gave written consent to participate in a crossover randomized controlled trial; all patients were subjected to both Macintosh and AirTraq laryngoscopy with manual inline stabilization after induction of anesthesia. The C-spine motion was examined at four areas: the occiput-C1 junction, C1-C2 junction, C2-C5 motion segment, and C5-thoracic motion segment. The time required for laryngoscopy was also measured.

Results: C-spine motion using the AirTraq was less than that during Macintosh laryngoscopy, averaging 66% less (P < 0.01) at three of the motion segments studied, occiput-C1, C2-C5, and C5-thoracic. There was no difference at the C1-C2 segment. There was no significant difference in the time to accomplish laryngoscopy between the two devices.

Conclusions: For patients in whom C-spine movement is undesirable, use of the AirTraq Laryngoscope may be useful to limit movement without an increase in the duration of intubation.

PubMed Disclaimer

Publication types

MeSH terms